Cook IVC Filter Study
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 473
- 试验地点
- 56
- 主要终点
- The Number of Participants With Technical Placement Success and With 12-month Freedom From New Symptomatic Pulmonary Embolism (PE) While a Filter is Indwelling.
研究概览
简要总结
This prospective, multicenter, double-arm clinical study further evaluated the safety and effectiveness of Cook's commercially available inferior vena cava (IVC) filters (specifically, the Günther Tulip® Vena Cava Filter and Cook Celect® Vena Cava Filters) in patients in need of temporary or permanent IVC filter placement for the prevention of pulmonary embolism (PE).
研究设计
- 研究类型
- Interventional
- 分配方式
- Non Randomized
- 干预模型
- Parallel
- 主要目的
- Treatment
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Requires temporary or permanent IVC filter placement for the prevention of PE
排除标准
- •Less than 18 years of age
- •Known hypersensitivity or contraindication to contrast medium that cannot be adequately premedicated
- •Known allergy or sensitivity to cobalt, chromium, or nickel
- •Pregnant or planning to become pregnant in the next 12 months
- •Patient refuses blood transfusions
- •At risk of septic embolism
- •Medical condition or disorder that would limit life expectancy to less than 12 months or that may cause noncompliance with the protocol or confound the data analysis
- •Existing IVC filter
- •Duplicate IVC
- •Anatomy that would prevent safe filter placement (e.g., condition of access vessels)
- •IVC diameter > 30 mm or < 15 mm
结局指标
主要结局
The Number of Participants With Technical Placement Success and With 12-month Freedom From New Symptomatic Pulmonary Embolism (PE) While a Filter is Indwelling.
时间窗: 12 months
The protocol defined technical placement success as: deployment of filter in a location suitable to provide sufficient mechanical protection against PE with no filter deformation, fracture, premature release, or clinical migration (caudal or cranial movement of a filter resulting in a surgical or endovascular intervention).
The Rate of 12-month Freedom From Major Adverse Events (MAEs)
时间窗: 12 months
MAEs were defined as clinical perforation, clinical migration, clinical fracture, embolization of the filter or filter fragments to the heart or lungs, IVC thrombotic occlusion, new symptomatic DVT while the filter is indwelling, access site complications with clinical sequelae, and procedure-/device-related death. Clinical perforation - Protrusion of filter legs through the wall of the IVC causing hemorrhage or hematoma or touching, impressing, or perforating another organ. Documented using CT. Clinical migration - Caudal or cranial movement of a filter resulting in surgical or endovascular intervention. Clinical fracture - A loss of structural integrity (breakage or separation) of the filter identified by imaging and associated with clinical sequelae and/or requiring intervention.
Supplemental Analysis to the Safety Endpoint: The Number of Participants With 12-month Freedom From Major Adverse Events (MAEs).
时间窗: 12 months
MAEs were defined as clinical perforation, clinical migration, clinical fracture, embolization of the filter or filter fragments to the heart or lungs, IVC thrombotic occlusion, new symptomatic DVT while the filter is indwelling, access site complications with clinical sequelae, and procedure-/device-related death. A filter retrieval is considered a successful safety result through 12 months, regardless of when the retrieval occurred.
次要结局
- Rate of 12-month Freedom From Major Adverse Events for the Combined Patient Population (i.e., the Celect and Günther Tulip Strata)(12 months)
- Number of Participants With Freedom From Grade 2 or Grade 3 Filter Leg Interaction With IVC, Filter Migration, Filter Fracture, and Filter Embolization at 12-month.(12 months)
- Rate of Technical Placement Success and 12-month Freedom From New Symptomatic PE While Filter is Indwelling(12 months)