Validation of the BC1 Bioelectrical Impedance Body Comp Analyzer
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 362
- 试验地点
- 2
- 主要终点
- Percent Body Fat
研究概览
简要总结
The purpose of this study was to validate the Stayhealthy BC1 bioelectrical impedance analyzer to assess percent body fat in children and adults. The BC1 was compared to dual-energy x-ray absorptiometry assessment of body fat for adults and to hydrostatic weighing assessment of body fat for children. We hypothesized that the BC1 would provide similar measures of body fat between devices with acceptable limits of agreement.
详细描述
The BC1 body composition analyzer is manufactured by Stayhealthy. It is a handheld, noninvasive bioelectrical impedance device. When held, the BC1 sends an electrical impulse through the body to determine body fat, muscle mass, and body hydration. Bio-impedance devices measure the change in electrical resistance in body tissues by using a detectable electrical signal that passes through the body. The method is based on the principle that lean body mass contains virtually all the water and conducting electrolytes in the body and provides a good electrical pathway. In contrast, fat or fat-containing tissues produce a poor electrical pathway. By inducing a low-energy, high -frequency signal, a measurement of the baseline resistance to flow of electrical current can be assessed. The resistance measurement relates directly to the volume of the conductor. Through repeated measurements and identifying other parameters (i.e. gender, weight, age and height) proprietary equations were developed that determine total body water, fat free body mass, and total body fat.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Health Services Research
- 盲法
- None
入排标准
- 年龄范围
- 10 Years 至 80 Years(Child, Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •boys and girls 10-17 yrs and men and women 18-80 yrs
排除标准
- •acute illness
- •pregnancy
- •pacemaker or defibrillator
结局指标
主要结局
Percent Body Fat
时间窗: 2 measures were assessed during a single visit
Percent body was compared between devices
次要结局
未报告次要终点
研究者
Todd Schroeder
Associate Professor
University of Southern California