Repeatability and Reproducibility of the CADence™System
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 79
- 试验地点
- 3
- 主要终点
- Positive Percent Agreement and Negative Percent Agreement
研究概览
简要总结
This study is designed as a prospective study to capture data for evaluating the repeatability and reproducibility of the CADence™ System. The study will be conducted at 3 different study sites using 6 CADence™ systems. Each subject will be tested 4 separate times by 2 different operators at each site and will be tested at all 3 site locations in order to accomplish this evaluation. The study will enroll a total of at least 24 CADence positive patients and at least 24 CADence negative patients.
详细描述
The CADence™ system is comprised of a digital stethoscope used to record cardiac (heart) sounds, with integrated sensor used to record electrical activity of the heart (ECG), a single-use patient booklet, and the CADence™ Software application. CADence™ Software is a clinical decision support tool for the noninvasive functional evaluation of patients presenting with chest pain and at least two coronary artery disease risk factors that is intended to aid a qualified clinician's analysis of normal/physiological and pathological heart murmurs and in ruling out significant coronary disease as a cause for these symptoms of cardiovascular disease, utilizing simultaneous recording of cardiac sounds and ECG.
The purpose of this study is to assess the precision of the CADence site using multiple operators, cadence systems and clinical sites.
研究设计
- 研究类型
- Observational
- 观察模型
- Case Control
- 时间视角
- Prospective
入排标准
- 年龄范围
- 22 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Subjects clinically indicated for nuclear stress test and who satisfy all the inclusion and exclusion criteria are eligible for participation.
- •Inclusion criteria:
- •Age 22 years or older
- •Willing and able to give informed consent
- •Clinical indication for nuclear stress test evaluation
- •Chest pain syndrome
- •Two or more coronary artery disease risk factors as defined by:
- •Dyslipidemia: LDL >130 mm/dL or HDL <35 mm/dL or on treatment for dyslipidemia
- •Hypertension: blood pressure >140 mmHg systolic, >90 mmHg diastolic or on blood pressure altering treatment
- •Obesity: BMI>28
- •Current cigarette smoking
- •Diabetes: Type 1 or 2
- •Family history: coronary disease in a first or second degree relative
排除标准
- •Subjects clinically indicated for nuclear stress test and who satisfy all the inclusion and exclusion criteria are eligible for participation.
- •Inclusion criteria:
- •Age 22 years or older
- •Willing and able to give informed consent
- •Clinical indication for nuclear stress test evaluation
- •Chest pain syndrome
- •Two or more coronary artery disease risk factors as defined by:
- •Dyslipidemia: LDL >130 mm/dL or HDL <35 mm/dL or on treatment for dyslipidemia
- •Hypertension: blood pressure >140 mmHg systolic, >90 mmHg diastolic or on blood pressure altering treatment
- •Obesity: BMI>28
- •Current cigarette smoking
- •Diabetes: Type 1 or 2
- •Family history: coronary disease in a first or second degree relative
- •Exclusion criteria:
- •Body Mass Index (BMI)<18.5 or BMI >40
- •Known coronary disease as defined as:
- •Prior bypass surgery or coronary stenting
- •Q-wave infarction on a past EKG (>0.01 sec Q-wave duration in two adjacent leads)
- •Presence of pacemaker/defibrillator
- •Presence of artificial valve
- •Presence of obvious cyanotic or pre-diagnosed congenital heart defect and coarctation of the aorta
- •The presence of murmurs including moderate to severe valve disease, ventricular septal defects, and AV fistulae
- •Presence of moderate or severe valve disease as defined by >1/6 heart murmur on physical diagnosis
- •Left Ventricular Assist Device (LVAD)
- •Presence of scars on the site thorax areas
- •Participation in trial within 30 days prior to collecting CADenceTM data except participation in registry studies
- •Asthma with wheezing
- •Inability to lie flat in the supine position
- •Acute coronary syndrome with elevated cardiac biomarkers (TP>3x upper limit of normal (ULN) or CKMB>3x ULN)
- •Heart transplant
- •Current cocaine use (within the past 24 hours, as reported by subject)
- •Chronic Obstructive Pulmonary Disease (COPD)
- •Patient is not an acceptable candidate for CT angiography:
- •Renal failure with GFR <50
- •Iodinated contrast allergy
- •Elevated heart rate which cannot be controlled sufficiently to achieve a good CT angiogram
- •Body weight >350lbs
- •Persistent sinus rhythm abnormalities (e.g., atrial fibrillation) based on medical history.
结局指标
主要结局
Positive Percent Agreement and Negative Percent Agreement
时间窗: 14 days
repeatability and reproducibility
次要结局
未报告次要终点