Multicenter Registry for S-ICD Electrodes
试验速览
- 阶段
- 不适用
- 状态
- 尚未招募
- 入组人数
- 1,000
- 试验地点
- 1
- 主要终点
- Sense B Electrode Failure
研究概览
简要总结
The subcutaneous implantable cardioverter defibrillator (S-ICD) is established in the prevention of sudden cardiac death, but recently a defect of the electrode's Sense-B component was reported in some patients. This lead to inadequate shocks due to noise. The cause and incidence of this defect are unknown. The manufacturer issued a warning regarding this defect and a possible premature depletion of the battery in some device. Independent data on the incidence of these issues is limited. Further data is needed to better understand the risk of these complications.
详细描述
The S-ICD therapy is an established treatment for primary and secondary prevention of sudden cardiac death. Recently, a report was published about a previously described complication of the electrode, in which a defect of the so-called Sense-B component is reported in a few affected patients. The defect of this component of the electrode leads to electrode noise and subsequent inadequate shocks. The cause of the defect and the incidence are currently unknown. The manufacturer Boston Scientific has issued a warning and advisory.
In addition to this advisory regarding the electrode of the S-ICD, there is also an active warning from the manufacturer regarding the possible premature exhaustion of the S-ICD battery. This is caused by an error in the manufacturing process of approximately 40,000 devices produced in 2018. The manufacturer estimates the incidence of premature battery exhaustion to be 4-15% within 5 years. Independent data on the incidence of premature exhaustion of these devices were initially scarce. The investigators conducted a retrospective data query in 2021 to capture the incidence of battery exhaustion at that time. The incidence was 3.7% after 5 years. The average follow-up time was only 3 years. The incidence of premature exhaustion increased significantly after 3 years, so data with a longer follow-up period is urgently needed.
Given that these data already show that the incidence of premature exhaustion increases significantly after 3 years and the observation period in the retrospective data was on average slightly under 3 years, a further query of the existing clinical data on the battery as part of this register is also planned. This would allow for a more accurate estimate of the risk of premature battery depletion.
研究设计
- 研究类型
- Observational
- 观察模型
- Cohort
- 时间视角
- Retrospective
入排标准
- 年龄范围
- 8 Years 至 —(Child, Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •status post S-ICD insertion
排除标准
- •no available follow-up data beyond 2 weeks after insertion
结局指标
主要结局
Sense B Electrode Failure
时间窗: 3 months to 70 months after device insertion
Incidence of the Sense B Electrode Failure of the S-ICD Lead.
次要结局
- Premature Battery Depletion(3 months to 70 months after device insertion)