Incidence & Predictors of Mechanical Abnormalities of QuickSite® and QuickFlex® Left Ventricular Pacing Leads: Impact on Electrophysiological Profile

Dhanunjaya Lakkireddy, MD, FACC1 个研究点 分布在 1 个国家目标入组 190 人开始时间: 2012年7月最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
190
试验地点
1
主要终点
Evaluate the mechanical/structural abnormalities in implanted St Jude Medical QuickSite® and QuickFlex® left ventricular (LV) pacing leads

研究概览

简要总结

On April 2, 2012 the manufacturer of the Quicksite® and Quickflex® leads, St. Jude Medical, released a voluntary medical device advisory regarding these leads' potential for problems with the silicone insulation covering one end of the lead, where it meets another section of the lead which is coated with a different insulating material (polyurethane). There has been a voluntary suspension of sales of these leads.

There are approximately 179,000 of the above leads implanted worldwide with an estimated 101,000 leads currently in service (59,000 in US). There have been 39 confirmed cases of the conductor portion of the lead becoming exposed from the silicone insulation (externalization), and the rate of occurrence of this type of problem is conservatively estimated at less than one percent. However, a review done by St. Jude Medical of returned Quicksite® and Quickflex® leads and available X-ray and fluoroscopic (x-ray movie) images of patients with these leads (from 1,219 patients) revealed a higher rate of 3-4%. It is presumed that there are patients whose leads have not yet been affected by this problem, but whose leads may show signs of structural changes on x-ray or fluoroscopy. This study seeks to determine what structural changes may be occurring in these leads, what signs might predict these changes, and what impact these changes may have on the lead and its performance.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Patient that received Quicksite or Quickflex lead at the University of Kansas Hospital

排除标准

  • Not willing to consent to data collection

结局指标

主要结局

Evaluate the mechanical/structural abnormalities in implanted St Jude Medical QuickSite® and QuickFlex® left ventricular (LV) pacing leads

次要结局

未报告次要终点

研究者

发起方
Dhanunjaya Lakkireddy, MD, FACC
申办方类型
Other
责任方
Sponsor Investigator
主要研究者

Dhanunjaya Lakkireddy, MD, FACC

Associate Professor

University of Kansas Medical Center

研究点 (1)

Loading locations...

相似试验