跳至主要内容
临床试验/NCT02666222
NCT02666222已完成不适用

To Evaluate the Long-term Safety and Efficacy of the Esteem® Hearing Implant in Subjects Suffering From Moderate to Severe Hearing Loss

Envoy Medical Corporation5 个研究点 分布在 1 个国家目标入组 51 人开始时间: 2010年8月最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
51
试验地点
5
主要终点
Change From Baseline (Pre-implant Aided Condition) at Year 5 in Speech Reception Threshold (SRT)

研究概览

简要总结

The purpose of this study is to evaluate the long-term safety and efficacy of the Esteem® Hearing Implant in subjects suffering from moderate to severe hearing loss.

详细描述

The Post Approval Study is an extension of the pivotal clinical trial, which was designed as a prospective, multi-center, non-randomized, clinical trial to evaluate the safety and efficacy of the Esteem® System. For this trial the subject acts as his or her own control.

This trial has been designed to meet the United States' regulatory requirements.

The subjects included in this study include the 57 subjects implanted under the original Investigational Device Exemption (IDE) study and the five (5) subjects implanted under the continued access expansion approved by the FDA. Of these 62 subjects, all but one (explant) were available for continued follow-up. These subjects were re-consented for the 5 year Post Approval Study. No new subjects were enrolled and implanted in this study. The goal was to monitor a minimum of 45 subjects through their five (5) year follow-up. Ultimately, 51 subjects completed the study.

The following questions are to be answered:

  • Is the Esteem effective through 5 year follow-up?
  • Is the Esteem safe through 5 year follow-up?

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
18 Years 至 90 Years(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Subject must meet all of the following criteria to be eligible for treatment in the trial:
  • Subject is at least 18 years old
  • Subject understands the nature of the procedure and has signed the Subject Informed Consent Form prior to the procedure
  • Subject is willing and able to comply with specified follow-up evaluations and understands the audiological test procedures and use of the Esteem® System.
  • Subject has moderate to severe sensorineural hearing loss between 500 and 4000 Hz in the ear to be implanted with pure tone air-conduction threshold levels within the limits of a Hearing Aid (HA) as follow:
  • Freq (Hz) 500 1000 2000 3000 4000 Lower Level (dB HL) 30 35 35 35 35 Upper Level (dB HL) 100 100 100 100 100
  • Subject's air-bone gap is no greater than 10 dB at 4 of the 5 following frequencies: 500, 1000, 2000, 3000 and 4000 Hz.
  • Subject has an unaided maximum word recognition score of greater than or equal to 40% with recorded delivery using a phonetically balanced word list at Speech Reception Threshold (SRT) + 40 dB or at maximum tolerable presentation level.
  • Subject is a current user of a properly functioning and appropriately fit hearing aid. This is defined as the subject has used this aid for at least four (4) hours (average) per day (in the ear to be implanted) for at least three (3) months for a new aid or one (1) month for an adjusted aid.
  • Subject's hearing aid, in the ear to be implanted, shall appropriately fit optimally.
  • Subject has normally functioning eustachian tube
  • Subject has normal tympanic membrane
  • Subject has a normal middle ear anatomy
  • Subject has adequate space for Esteem® System implant determined via fine cut temporal bone CT scan
  • Subject is a native speaker of the English language.
  • Subject is a hearing aid user in the ear to be implanted

排除标准

  • Subjects will be excluded from the trial if any one of the following criteria is met:
  • Subject has a history of post-adolescent chronic middle ear infections, inner ear disorders or recurring vertigo requiring treatment, disorders such as mastoiditis, Hydrops or Meniere's syndrome or disease
  • Subject has a history of external otitis or eczema for the outer ear canal and the investigator believes this will affect the Esteem® System implantation
  • Subject has cholesteatoma or destructive middle ear disease
  • Subject has life expectancy of less than two (2) years due to other medical conditions
  • Subject has retrocochlear or central auditory disorders
  • Subject is known to be suffering from any psychological, developmental, physical, or emotional disorder that the investigator feels would interfere with the surgery or follow-up testing
  • Subject has a known history of fluctuating air conduction and/or bone conduction hearing loss over a one-year period of 15 dB in either direction at 2 or more frequencies (from 500 - 4000 Hz)
  • Subject has sudden hearing loss due to unknown cause
  • Subject has a history of disabling tinnitus, defined as tinnitus which required treatment.
  • Subject is unable to adequately perform audiological testing
  • Subject has a medical condition or undergoing a treatment that may affect healing and the investigator does not believe the subject is a good candidate for the trial.
  • Subject has diabetes that is not well controlled with medication or diet and the investigator does not believe in his best medical judgment that the subject would be a good candidate for the trial
  • Subject is pregnant at the time of device implant
  • Subject has a history of keloid formation
  • Subject has known hypersensitivity to silicone rubber, polyurethane, stainless steel, titanium and/or gold

结局指标

主要结局

Change From Baseline (Pre-implant Aided Condition) at Year 5 in Speech Reception Threshold (SRT)

时间窗: Baseline through Year 5 of Follow Up

ENDPOINT #1: SRT at baseline minus SRT at Year 5. Positive difference (i.e. lower value of SRT with the Esteem) indicates better outcome.

Change From Baseline (Pre-implant Aided Condition) at Year 5 in Word Recognition Score (WRS) at 50 dB HL

时间窗: Baseline through Year 5 of Follow Up

ENDPOINT #2: WRS at Year 5 minus WRS at baseline. Positive difference (in % correct) indicates better outcome.

Incidence of Serious Adverse Device Events (SADEs) and Device Failures and Replacements at Each Follow-up.

时间窗: SADEs, PAS phase through Year 5 of Follow Up

ENDPOINT #3: The analysis of the incidence of SADEs and device failures and replacements at each follow-up.

Bone Conduction Stability

时间窗: Baseline through 5 Year Follow-Up

ENDPOINT #5: Difference between Baseline and 5 Year Pure-Tone Average (PTA; average of 500, 1000, 2000 Hz thresholds); calculated as PTA at Year 5 minus PTA at baseline. Smaller magnitude dB difference indicates better outcome.

次要结局

  • Improvement in Quality of Life as Reflected by Abbreviated Profile of Hearing Aid Benefit (APHAB) Questionnaire(Baseline through Year 5 of Follow Up)

研究者

申办方类型
Industry
责任方
Sponsor

研究点 (5)

Loading locations...

相似试验