The Vascutek Custom Fenestrated Anaconda Post-Market Surveillance Study

Vascutek Ltd.70 个研究点 分布在 8 个国家目标入组 628 人开始时间: 2010年6月最近更新:
适应症

试验速览

阶段
不适用
状态
终止
发起方
入组人数
628
试验地点
70
主要终点
To monitor the the post-market clinical safety and effectiveness of the Vascutek AnacondaTM Custom Fenestrated Stent Graft

研究概览

简要总结

This is a multicentre, prospective, observational post-market registry (follow-up 1 month, 1 year, 2 years, 3 years, 4 years and 5 years) to monitor the post-market clinical safety and effectiveness of the Vascutek AnacondaTM Custom Fenestrated Stent Graft in patients with a short infrarenal neck of <15mm or juxta-renal abdominal aortic aneurysms who, in the opinion of the treating physician, are anatomically suitable for the AnacondaTM Custom Fenestrated Stent Graft and for whom a customised medical device is required.

详细描述

AnacondaTM Custom Fenestrated Devices (CFD) are designed for the endovascular repair of abdominal aortic aneurysms (AAA) for cases where the aneurysm extends superiorly very close to or involving the renal arteries, leaving insufficient neck length for infrarenal device sealing. The non-CE marked devices are custom designed based on CT scan data of the patient's aorta and regulated in accordance with the Medical Device Directive (93/42/EEC) and associated country specific regulations as applicable.

The proximal end of the AnacondaTM fenestrated body device is positioned suprarenally, where it seals and fixates in the vessel by means of rings comprised of multiple turns of fine nitinol wire, attached to the fabric with polyester sutures, creating a self-expanding stent which provides a conformable seal with the vessel wall and assists in anchoring of the device. Nitinol hooks provide secure fixation within the suprarenal aorta. The saddle shape of the rings allows the superior mesenteric artery (SMA) or celiac artery to be accommodated in the valley of the top ring saddle in some cases.

The objective of the registry is to provide long term (5 year) safety and effectiveness data of all patients who consent to their data to be used in the registry. This data aims to demonstrate that the Vascutek AnacondaTM Custom Fenestrated Stent Graft is safe and effective in patients with a short infrarenal neck of <15mm or juxta-renal abdominal aortic aneurysms who, in the opinion of the treating physician, are anatomically suitable for the AnacondaTM Custom Fenestrated Stent Graft.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • A patient with juxta renal abdominal aortic aneurysms (AAA) who, in the opinion of the treating physician, is anatomically suitable for the Anaconda Fenestrated Graft and for whom a customised medical device is required.

排除标准

  • Ruptured or symptomatic aneurysm
  • Clinically significant concomitant medical disease or infection
  • Connective Tissue Disease (e.g. Marfan's Syndrome)
  • Known allergy to nitinol, polyester or contrast medium
  • Excessive tortuosity of access vessels (femoral or iliac arteries)
  • Landing zone of less than 10mm in the visceral segment of the aorta
  • Inability to comply to follow up protocol
  • Access vessels less than 6mm in diameter
  • Diseased or excessively tortuous access to target vessels
  • Target vessels of less than 5mm in calibre
  • Excessive calcification or thrombus at the intended landing zone which could affect sealing

结局指标

主要结局

To monitor the the post-market clinical safety and effectiveness of the Vascutek AnacondaTM Custom Fenestrated Stent Graft

时间窗: For 5 years post-implant

To monitor the post-market clinical safety and effectiveness of the Vascutek AnacondaTM Custom Fenestrated Stent Graft in patients with a short infrarenal neck of \<15mm or juxta-renal abdominal aortic aneurysms who, in the opinion of the treating physician, are anatomically suitable for the AnacondaTM Custom Fenestrated Stent Graft and for whom a customised medical device is required.

次要结局

未报告次要终点