In vivO Study of assessmeNt for Efficacy and safetY of Zotarolimus-eluting Stent OVERexpansion With Optical Coherence Tomography (ONYSOVER Trial)

Yonsei University2 个研究点 分布在 1 个国家目标入组 30 人开始时间: 2022年3月23日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
进行中(未招募)
入组人数
30
试验地点
2
主要终点
Maximal stent overexpansion (%)

研究概览

简要总结

Investigators aimed to evaluate efficacy and safety of expansion capacity of zotarolimus-eluting Stent assessed by optical coherence tomography (OCT) in vivo study.

详细描述

Single stent implantation (also called "provisional stenting") is currently the preferred technique for percutaneous coronary interventions on the majority of unprotected left main (LM) or non-LM bifurcations. During provisional stenting, proximal optimization technique (POT), proved to correct both the proximal main vessel (MV) malapposition and/or optimize the side branch (SB) ostium strut opening for proper wire re-crossing into the distal cell, might be crucial step and may warrant improved long-term clinical outcomes in the setting of major proximal and distal reference vessel size mismatch (like often in LM bifurcation or diffuse tapered long lesion).

Recently, second-generation drug-eluting stent (DES) implantation is recommended over BMS for PCI in patients with ischemic heart disease (IHD) due to superior efficacy and safety. However, percutaneous coronary intervention (PCI) with DES for lesion with the discrepancy of proximal and distal reference vessel size, such as LM or non-LM bifurcation lesion, diffuse tapered long lesion, and ectatic or aneurysmal coronary artery, is challenging due to stent optimization because expert consensus recommend that stent diameter is selected by distal reference diameter, which need stent overexpansion for POT.

OCT is a utility that can accurately measure reference vessel diameter and lesion length pre-PCI, which is useful for stent optimization by a suitable stent selection and pre- and post-interventional strategies. OCT also demonstrated higher sensitivity to detect stent malapposition, edge dissection, and tissue prolapsed than IVUS post-PCI. Thus, OCT-guided PCI can be helpful to optimize DES implantation in patients with dilated coronary arteries despite the potential limitation of depth penetration and attenuation images by thrombus.

Zotarolimus-eluting stent (Onyx family stent, Medtronic, Santa Rosa, CA, USA) was developed in response to the demand for stents with improved radiographic visibility. It has a novel thin strut composite wire stent platform that is covered with the same zotarolimus-eluting durable polymer coating as its predecessors. The metallic stent platform consists of a composite wire made from a dense platinum-iridium core, which makes the struts radiopaque, and an outer layer of cobalt-chromium alloy. The dense core also allows for reduced strut thickness, which might be associated with a decreased risk of stent thrombosis.

Onyx family stent has been labeled for maximum stent inner diameter by easy crowns platform (2.00-2.5 mm, 6.5 Crowns; 2.75-3.0 mm, 8.5 crowns; 3.5-4.0 mm, 9.5 crowns; 4.5-5.0 mm; 10.5 crowns). However, safety and efficacy of DES overexpansion is still concerned as safety of efficacy of there are limited data regarding Onyx family stent overexpansion in vivo.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
19 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Patients, ≥19 years of age, who were diagnosed with IHD requiring stent implantation for lesion with size discrepancy between proximal and distal reference diameter over 1mm, such as tapered long lesion or bifurcation lesion, confirmed by OCT
  • The decision to participate voluntarily in this study and the written consent of the patient

排除标准

  • Patients with hypersensitivity or contraindication to antiplatelet treatment
  • Female of childbearing potential, who possibly plans to become pregnant any time after enrollment into this study.
  • Patients with a life expectancy shorter than 1 year

研究组 & 干预措施

In vivo

OCT examination is performed using a frequency-domain OCT system (C7 DragonflyTM OPTISTM Imaging Catheter, Abbott Vascular, Santa Clara, CA, USA) with a 5.4 or 7.5 cm total pullback length according to a non-occlusive technique. The procedure is performed via radial or femoral access with a ≥6 Fr guiding catheter. The OCT catheter is advanced over the 0.014-inch PCI wire and the implanted stent was crossed after administration of intracoronary nitrates of 200 µg. For blood clearing, contrast media was injected through the guiding catheter with an automated power injector. The standard infusion rate was 4 mL/s for 4 seconds with 250 PSI. Follow-up OCT image acquisition was achieved with the same method. After post-PCI OCT, we evaluate the factor regarding post-PCI optimization target by European Expert Consensus.

干预措施: Optical coherence tomography-guided percutaneous coronary intervention (Device)

In vivo

OCT examination is performed using a frequency-domain OCT system (C7 DragonflyTM OPTISTM Imaging Catheter, Abbott Vascular, Santa Clara, CA, USA) with a 5.4 or 7.5 cm total pullback length according to a non-occlusive technique. The procedure is performed via radial or femoral access with a ≥6 Fr guiding catheter. The OCT catheter is advanced over the 0.014-inch PCI wire and the implanted stent was crossed after administration of intracoronary nitrates of 200 µg. For blood clearing, contrast media was injected through the guiding catheter with an automated power injector. The standard infusion rate was 4 mL/s for 4 seconds with 250 PSI. Follow-up OCT image acquisition was achieved with the same method. After post-PCI OCT, we evaluate the factor regarding post-PCI optimization target by European Expert Consensus.

干预措施: Optical coherence tomography-guided Onyx family stent expansion (Device)

结局指标

主要结局

Maximal stent overexpansion (%)

时间窗: Through procedure completion, up to 24 hours

defines as the ratio of the maximal stent diameter to each diameter of stent

次要结局

  • Grade of polymer damage assessed by electron microscope(Through procedure completion, up to 1-month)
  • Absolute average stent diameter (mm) in the maximal stent expansion site(Through procedure completion, up to 24 hours)
  • Stent strut fracture(Through procedure completion, up to 24 hours)

研究者

申办方类型
Other
责任方
Principal Investigator
主要研究者

Yongcheol Kim

Clinical Associate Professor

Yonsei University