Evaluation of a Safe Infusion Device in Reducing Occupational Exposure of Nurses to Antineoplastic Drugs: a Comparative Prospective Study. CONTAMOINS-1
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 120
- 试验地点
- 1
- 主要终点
- Rate of contamination of nurses' gloves with cytotoxic products
研究概览
简要总结
Monocentric comparative before / after study to evaluate the efficacy of Safe Infusion Devices (SIDs) in reducing drug exposure to nurses compared to usual perfusion practices.
详细描述
Despite the decreasing of environmental contamination throughout the anticancer drug circuit, the administration of chemotherapies remains at risk of occupational exposure for nurses. Many medical devices aim at securing administration, but none have been scientifically evaluated to verify the actual improvement.
A monocentric comparative before / after study was carried out in an oncology day hospital in order to evaluate the efficacy of Safe Infusion Devices in reducing drug exposure compared to usual perfusion practices (neutral solvent-purged infusers). The rate of nurses' gloves contamination was estimated. The eligible protocols must contain at least one disconnection step and one of the 10 drugs screened. To avoid false negatives, each sample of gloves was contaminated with a drop of topotecan. Association between contamination and other variables was investigated using a multivariate regression analysis.
The usual practice leaded to a rate of 58.3% of contaminated samples while Safe Infusion Devices to a rate of 15%: Safe Infusion Devices reduced the risk of gloves contamination by 84% in multivariate analysis (Odds ratio=0.16; 95% confidence interval=0.05-0.47; p<0.001). Topotecan was identified within 100% of the samples. Only one case of cross-contamination has occurred.
Despite the current practice of using neutral solvent-purged infusers, the occupational exposure remains high for nurses and Safe Infusion Devices significantly reduced this risk of exposure. However, glove contamination is only a surrogate endpoint. The results confirmed that the disconnection of empty bags resulted in occupational exposure. Except a contamination due to the leakage of a bag, no cross-contamination was detected. This validated the environmental quality of the cytotoxic drug circuit. Safe Infusion Devices were highly effective but did not completely eliminate exposure.
研究设计
- 研究类型
- Observational
- 观察模型
- Case Control
- 时间视角
- Prospective
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Protocols which contained at least one of the 10 drugs of the dosing assay
- •Protocols with at least one disconnection step.
- •Given the large number of eligible protocols, we favored those with the most concentrated products (CPM, PMX) to maximize the chance of detecting contamination and also to extend inclusions over a longer period.
排除标准
- •Drugs administrated with syringes (intravenously or subcutaneously) were not included because of the absence of impact of the SIDs on these modalities of administration.
结局指标
主要结局
Rate of contamination of nurses' gloves with cytotoxic products
时间窗: day 1
The statistical unit consists of a sample taken from all the gloves used for the administration of a daily course. The dosing assay allowed to identify a panel of 10 cytotoxic drugs (HPLC MS/MS).
次要结局
- Costs(day 1)
- Quantitative contamination of nurses' gloves with cytotoxic products(day 1)
- Manufacturing time(day 1)