Optical Imaging Measurement of Intravascular Solution Efficacy Trial
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 23
- 试验地点
- 2
- 主要终点
- Quality of Images
研究概览
简要总结
Iodinated contrast is the current gold standard for infrainguinal angiography imaging in patients without renal insufficiency and has also been used with intravascular Optical Coherence Tomography (iOCT) to improve image quality in human coronary arteries as well as carotid arteries. The current debate in the literature for iOCT medium is between iodinated contrast and dextran and CO2 may offer a superior method of iOCT imaging during lower extremity occlusive disease interventions.
The investigators hypothesize that the CO2 medium injection during iOCT data acquisition is feasible and will produce at least the same quality of imaging as that obtained with contrast or dextran without causing the problems of volume overload and renal toxicity seen with the two latter mediums.
Primary Outcomes Measured
- Quality: Cumulative number of clear image frame (CIF) through the entire 54mm length segment.
- Quantitative: Calculations of the area and diameter of each segment will be measured to determine if index of refraction has any effect between the three mediums to be tested.
The investigators expect to find little difference between all three iOCT mediums and hope to conclude that CO2 offers a superior side effect profile for iOCT imaging in the lower extremity arterial system.
详细描述
Peripheral artery disease (PAD) affects anywhere from 8-12 Million people in the United States. Many of these people go on to develop claudication, rest pain, and tissue loss. During the workup for these disease states many imaging modalities are conducted including Pulse Volume Recording, Duplex Ultrasound, Angiography, and IVUS, but an emerging catheter based imaging has been developed that may supplement the current modalities used.
Intravascular optical coherence tomography (iOCT) is based on near-infrared light system. The light reflects off plaque and other objects within vessels and the signals are processed into a series of axial images (A-scans) at different positions along the artery to generate a two-dimensional dataset (B-scans). These images are created at an extremely fine resolution of 10-15 μm, which has allowed iOCT to be used in many research settings including PAD and coronary artery disease. OCT has been approved for clinical use in the coronary territory by the FDA in May 2010. Since then many centers have been using iOCT in the daily clinical practice. However, it's still not widely in the clinical management of patients with PAD. There is hope that the high resolution capabilities of iOCT may help before and after an intervention to predict outcomes or correct errors in stent deployment.
The iOCT procedure for lower extremity PAD is fairly straightforward. An introducer is placed into the femoral artery. After which a wire is placed past the lesion of interest and the iOCT catheter is inserted. The catheter is then attached to an automated pullback device. Next an optical medium is needed to displace the erythrocytes. Due to the high resolution of the iOCT this is necessary for a cleaner image to analyze. At the time of injection of the optical medium a sensor triggers the catheter to be withdrawn (distal to proximal) at anywhere from 10-25 mm/sec. The images are captured and processed and arterial plaque can be characterized.
The greatest strength of the iOCT catheter is its high resolution images but the problem is that the imaging signal is substantially attenuated by blood. In order to remedy this complication techniques such as proximal balloon occlusion and continuous infusion of a fluid have been used to acquire improved iOCT images. In the continuous infusion methods, different mediums have been injected such as contrast, dextran, and even an oxygen-carrying substitute in hopes of improving decreasing the attenuation by blood.
In order the overcome the attenuation of red blood cells during iOCT imaging, we are proposing a novel approach involving CO2 injection to clear the erythrocytes. Currently CO2 is used as medium for digital subtraction angiography in patients with renal insufficiency and was first used in humans by Hawkins in 1982. The other alternative for angiography, iodinated contrast medium, is nephrotoxic and thus is avoided in these patients for fear of exacerbating the patient's acute or chronic problem. Another group where CO2 angiography is employed is history of a contrast allergy. Although this technique is usually used under these circumstances, Kerns et al reports conducting CO2 angiography in as high as 20% of their patients with abdominal and lower extremity studies.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Diagnostic
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •Age greater than or equal to 18 years
- •English speaking
- •Scheduled to undergo an infrainguinal angiogram and/or endovascular procedure as determined by a vascular surgery specialist
- •Superficial Femoral Artery diseased segment
排除标准
- •Acute or Chronic Renal insufficiency with Cr >1.5
- •Chronic obstructive pulmonary disease
- •Congestive heart failure (American Heart Association C lass III or IV)
- •Acute limb ischemia, defined by a significant change in symptoms (one category on the Rutherford scale within the previous 14 days)
- •Concurrent oral anticoagulant therapy that cannot be safely withheld
结局指标
主要结局
Quality of Images
时间窗: 1 month
The metric of image quality was the clear imaging field (CIF), which was defined as a cross section in which ≥270° of the vessel wall architecture was visualized. This has been used previously to quantify adequacy of clearance in OCT image comparison. Two independent observers, blinded to the flush medium used, analyzed all OCT frames in each pullback sequence. Any disagreement \>10% was resolved with a consensus re-evaluation at a later time point by the same reviewers. Each individual cross section was assigned a designation of quality or insufficient quality; thus, a quality image proportion was generated for each run by taking the mean of each observer's determinations
次要结局
- Superficial Femoral Artery Plaque Composition by Flush Medium(1 month)