Validation of Diagnostics to Identify Glucose-6-phosphate Dehydrogenase Activity in the US

PATH2 个研究点 分布在 1 个国家目标入组 250 人开始时间: 2019年5月13日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
250
试验地点
2
主要终点
Specificity of SD Biosensor POC G6PD Test for Identifying G6PD Deficient Individuals

研究概览

简要总结

The primary objective of this study is to assess the accuracy of the SD Biosensor STANDARD G6PD Analyzer in measuring G6PD activity when used by trained health care workers.

详细描述

This is a cross-sectional diagnostic accuracy study with up to 250 participants with a goal of obtaining 20 deficient and 20 intermediate samples. The clinic will recruit and consent study participants. Clinic staff will draw 3 whole blood samples and obtain finger stick capillary blood.

G6PD activity is reported in terms of grams of hemoglobin (Hb), hence the hemoglobin concentration must be measured.

Clinic staff will perform the investigational Standard Diagnostics (SD) Biosensor point-of-care (POC) test for glucose-6-phosphate dehydrogenase (G6PD) deficiency and a HemoCue® hemoglobin test on finger stick capillary blood and on the venous blood samples.

Another venous blood sample will be sent to a clinical laboratory improvement amendments (CLIA)-certified laboratory for reference testing by the gold standard assays:

  • G6PD measurement by spectrophotometry using the Pointe Scientific G6PD reference assay
  • hemoglobin measurement by a hematology analyzer

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Diagnostic
盲法
None

入排标准

年龄范围
18 Years 至 65 Years(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Willingness to provide consent

排除标准

  • Blood transfusion in the past 90 days by self-report

结局指标

主要结局

Specificity of SD Biosensor POC G6PD Test for Identifying G6PD Deficient Individuals

时间窗: All samples were collected on study day 1

For the purposes of this study, an individual was considered G6PD deficient if they tested positive by the Pointe Scientific assay. A true negative was defined as \> 30% of normal G6PD activity in circulating venous blood as determined by the Pointe Scientific test. Specificity is the percentage of participants who tested negative for G6PD deficiency on the reference test who were identified as negative using the test assay, calculated as: Number of participants who were both test and true negative / (Number of participants who were test and true negative + Number of participants who were test positive but true negative \[ie false positive\]) \* 100%. Specificity of the SD Biosensor G6PD test was calculated from both venous and capillary blood samples.

Sensitivity of SD Biosensor POC G6PD Test for Identifying G6PD Deficient Individuals

时间窗: All samples were collected on study day 1

For the purposes of this study, an individual was considered G6PD deficient if they tested positive in the Pointe Scientific assay. A true positive was defined as a participant with ≤ 30% of normal G6PD activity in circulating venous blood determined by the Pointe Scientific test. Diagnostic sensitivity of the SD Biosensor G6PD test is defined as the percentage of participants who tested positive for G6PD deficiency on the reference test who were identified by the test assay as positive, calculated as: Number of participants who were both test and true positive / (Number of participants who were test and true positive + Number of participants who were test negative but true positive \[ie false negative\]) \* 100%. Sensitivity of the SD Biosensor POC G6PD test was calculated from both venous and capillary blood samples.

Sensitivity of SD Biosensor POC G6PD Test for Identifying Women With Intermediate G6PD Activity

时间窗: All samples were collected on study day 1

To investigate the performance of the SD Biosensor G6PD test to distinguish females with intermediate G6PD activity from females with normal activity, assay sensitivity was determined at a G6PD activity threshold of 70%. A true positive for intermediate G6PD activity in females was defined as G6PD activity between 30 and 70% of normal in circulating venous blood as determined by the Pointe Scientific test. Sensitivity of the SD Biosensor POC G6PD test is the percentage of women who tested positive for intermediate G6PD activity on the reference test who were identified by the test assay as positive, calculated as: Number of women who were both test and true positive / (Number of women who were test and true positive + Number of women who were test negative but true positive \[ie false negative\]) \* 100%. Sensitivity of the SD Biosensor G6PD test for identifying females with intermediate G6PD activity was calculated from both venous and capillary blood samples.

Specificity of SD Biosensor G6PD Test for Identifying Women With Intermediate G6PD Activity

时间窗: All samples were collected on study day 1

To investigate the performance of the SD Biosensor G6PD test to distinguish females with intermediate G6PD activity from females with normal activity, the specificity was determined at a G6PD activity threshold of 70%. A true negative for intermediate G6PD activity in females was defined as G6PD activity \> 70% of normal in circulating venous blood as determined by the Pointe Scientific test. Specificity is the percentage of participants who tested negative for G6PD deficiency on the reference test who were identified as negative using the test assay, calculated as: Number of participants who were both test and true negative / (Number of participants who were test and true negative + Number of participants who were test positive but true negative \[ie false positive\]) \* 100%. Specificity of the SD Biosensor G6PD test was calculated from both venous and capillary blood samples.

次要结局

  • Accuracy Between the SD Biosensor POC G6PD Test Assay and the Pointe Scientific Test Kit(All samples were collected on study day 1)
  • Accuracy Between the SD Biosensor POC G6PD Test Measure of Hemoglobin and the Reference Hemoglobin Test(All samples were collected on study day 1)
  • Median G6PD Values Measured by the SD Biosensor G6PD Test for Venous and Capillary Blood Samples(All samples were collected on study day 1)