临床试验/NCT01510717
NCT01510717已完成不适用

Clinical Investigation of AcrySof® IQ ReSTOR® +2.5 D Multifocal Intraocular Lens (IOL) Model SN6AD2

Alcon Research0 个研究点目标入组 409 人开始时间: 2012年2月1日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
入组人数
409
主要终点
Photopic Contrast Sensitivity With Glare at Day 120-180

研究概览

简要总结

The purpose of this study is to compare the safety and efficacy of an investigational multifocal intraocular lens (IOL) to an FDA-approved monofocal IOL.

详细描述

Participants 21 years and older in need of cataract extraction were consented and randomized in a 1:1 ratio at each investigative site to receive either the AcrySof® IQ ReSTOR® +2.5 D Multifocal IOL Model SN6AD2 [SV25T0] or the AcrySof® IQ Monofocal IOL Model SN60WF in both eyes. Each participant completed a preoperative examination of both eyes, implantation of IOL at the operative visit for each eye, and up to 7 postoperative visits, with monocular testing at Day 1-2, Day 7-14, and Day 30-60, and monocular and binocular testing at Day 120-180 following the second implantation. The first eye to be implanted was considered the primary eye. The second eye was implanted 7 and 30 days of the first. If the first eye was not implanted during surgery, the second eye was not eligible. If the second eye was not implanted during surgery, only the first eye was followed.

研究设计

研究类型
Interventional
分配方式
Randomized
干预模型
Parallel
主要目的
Treatment
盲法
Double (Participant, Outcomes Assessor)

入排标准

年龄范围
21 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者
否

入选标准

  • •Diagnosis of bilateral cataracts
  • •Willing and able to complete all required postoperative visits
  • •Calculated lens power within the available supply range
  • •Planned cataract removal by phacoemulsification
  • •Potential postoperative visual acuity of 0.2 logMAR or better in both eyes
  • •Preoperative astigmatism less than 1.0 diopter
  • •Clear intraocular media other than cataract in study eyes
  • •Preoperative Best Corrected Distance Visual Acuity (BCDVA) worse than 0.2 logMAR
  • •Able to undergo second eye surgery between 7 days and 30 days of the first eye surgery
  • •Other protocol-defined inclusion criteria may apply

排除标准

  • •Previous refractive surgery
  • •Inflammation or edema of the cornea
  • •Optic nerve atrophy
  • •Pregnancy
  • •Current participation in another investigational drug or device study
  • •Other protocol-defined exclusion criteria may apply

研究组 & 干预措施

Multifocal IOL

Experimental

AcrySof® IQ ReSTOR® +2.5 D Multifocal IOL Model SN6AD2 [SV25T0], bilateral implantation

干预措施: AcrySof® IQ ReSTOR® +2.5 D Multifocal IOL Model SN6AD2 [SV25T0] (Device)

Monofocal IOL

Active Comparator

AcrySof® IQ Monofocal IOL Model SN60WF, bilateral implantation

干预措施: AcrySof® IQ Monofocal IOL Model SN60WF (Device)

结局指标

主要结局

Photopic Contrast Sensitivity With Glare at Day 120-180

时间窗: Day 120-180 from second eye implantation

Contrast sensitivity was assessed binocularly with the participant's best spectacle correction under photopic (bright) conditions at a distance of 8 feet at spatial frequencies of 3, 6, 12, and 18 cycles per degree (cpd) using the Vector Vision CSV 1000 with a glare source. Raw scores from contrast sensitivity testing were transformed to log units. Scores of (-1) were set to missing; hence, the mean measures may be overestimated and the variability measures may be underestimated. A higher numeric value represents better contrast sensitivity.

Mean Photopic Monocular Distance Corrected VA (53 cm) at Day 120-180

时间窗: Day 120-180 from second eye implantation

Visual acuity (VA) was tested monocularly (each eye separately) using the manifest refraction adjusted for optical infinity and the hand-held, 100% contrast, Early Treatment Diabetic Retinopathy Study (ETDRS) chart set at 53 centimeters (cm) on the nearpoint rod. VA was measured in logMAR (logarithm of the minimum angle of resolution), with 0.1 logMAR increment corresponding to 5 letters, or 1 line, on an ETDRS chart. A lower numeric value represents better visual acuity. This analysis was prespecified for the primary eye.

Number of Cumulative and Persistent Adverse Events as Defined in IS EN ISO 11979-7:2006, up to Day 120-180

时间窗: Day 0 first operative eye visit, up to Day 120-180 from second eye implantation

Cumulative and persistent adverse events were collected. This outcome measure was prespecified for the multifocal IOL.

Photopic Contrast Sensitivity Without Glare at Day 120-180

时间窗: Day 120-180 from second eye implantation

Contrast sensitivity (ie, the ability to detect objects by distinguishing them from their background) was assessed binocularly with the participant's best spectacle correction under photopic (bright) conditions at a distance of 8 feet at spatial frequencies of 3, 6, 12, and 18 cycles per degree (cpd) using the Vector Vision CSV 1000 without a glare source. Raw scores from contrast sensitivity testing were transformed to log units. Scores of (-1) were set to missing; hence, the mean measures may be overestimated and the variability measures may be underestimated. A higher numeric value represents better contrast sensitivity.

Mesopic Contrast Sensitivity Without Glare at Day 120-180

时间窗: Day 120-180 from second eye implantation

Contrast sensitivity was assessed binocularly with the participant's best spectacle correction under mesopic (dim lighting) conditions at a distance of 8 feet at spatial frequencies of 1.5, 3, 6, and 12 cycles per degree (cpd) using the Vector Vision CSV 1000 without a glare source. Raw scores from contrast sensitivity testing were transformed to log units. Scores of (-1) were set to missing; hence, the mean measures may be overestimated and the variability measures may be underestimated. A higher numeric value represents better contrast sensitivity.

Mesopic Contrast Sensitivity With Glare at Day 120-180

时间窗: Day 120-180 from second eye implantation

Contrast sensitivity was assessed binocularly with the participant's best spectacle correction under mesopic conditions at a distance of 8 feet at spatial frequencies of 1.5, 3, 6, and 12 cpd using the Vector Vision CSV 1000 with a glare source. Raw scores from contrast sensitivity testing were transformed to log units. Scores of (-1) were set to missing; hence, the mean measures may be overestimated and the variability measures may be underestimated. A higher numeric value represents better contrast sensitivity.

次要结局

  • Near Spectacle Independence Using SILVER Patient Reported Outcome (PRO) Questionnaire at Day 120-180(Day 120-180 from second eye implantation)
  • Mean Photopic Monocular Best Corrected Distance VA (4 m) at Day 120-180(Day 120-180 from second eye implantation)
  • Overall Spectacle Independence Using SILVER Patient Reported Outcome (PRO) Questionnaire at Day 120-180(Day 120-180 from second eye implantation)
  • Mean Photopic Monocular Distance Corrected Near VA at Standard Distance (40 cm) at Day 120-180(Day 120-180 from second eye implantation)

研究者

申办方类型
Industry
责任方
Sponsor