AI/ML Algorithm Training and Validation for the imPulse™ Tor System Self-Directed Systolic & Diastolic Blood Pressure Monitor.
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 15
- 试验地点
- 1
- 主要终点
- imPulse™ Tor System Blood Pressure Prediction
研究概览
简要总结
Self-measured, non-invasive accurate blood pressure monitoring continues to be a major challenge for automated vital sign measurement systems. The general notion is that with reliable, self-administered BP monitoring in the clinic and at home, health care providers will be able to diagnose hypertension among individuals at an early stage, including high risk patients in the community, and more quickly assess if prescribed treatment plans are working. The imPulse™ Tor System detects 1) audible and inaudible low-frequency, low-amplitude sounds generated by the body, including arterial pulse waveforms, and 2) ECG-derived heart cycle identification, which can be combined with the vibroacoustic data to estimate blood pressure. The imPulse™ Tor has undergone preliminary testing. In this pilot study, we collect data from health care workers for algorithm training and validation study to achieve medical grade device AAMI/ISO and IEEE standards compliance.
详细描述
Hypertension (HTN) is a major risk factor for cerebrovascular morbidity and mortality, yet its identification can be delayed due to lack of overt symptoms, relying on Blood Pressure (BP) measurements for diagnosis. Several BP monitoring techniques are used in clinics and hospitals, and there is also an outpatient method that is used for 24-hour BP monitoring, based on a sphygmomanometer. The general notion is that with frequent, reliable, self-directed BP monitoring in the clinic and at home, health care providers will be able to diagnose hypertension among individuals at an early stage, including high-risk patients in the community, and more quickly assess if prescribed treatment plans are working. However, non-invasive self-administered blood pressure monitoring continues to be a major challenge for automated vital sign measurement systems.
To circumvent the limitations of current systems of interval hemodynamic measurement and increasing demands on healthcare providers, a non-invasive automated self- directed vital sign monitor that can integrate into hospital early warning systems and warn healthcare providers of deteriorating vital sign parameters would be of significant value. This study aims to provide proof of concept for a portable device that measures vital signs (blood pressure, heart rate, and blood oxygen saturation) using biophysical pneumatics and hydraulics, vibroacoustics, and multi-lead ECG that aims to produce results that are as good as a sphygmomanometer-based device and pulse oximeter, without the need for a circumferential pressure device/sphygmomanometer on the arm.
研究设计
- 研究类型
- Observational
- 观察模型
- Cohort
- 时间视角
- Prospective
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •Participants are limited to clinical staff and first-responders who work at the UC Davis C Street Clinics in the Sports and Spine suites.
排除标准
- •History of major neck surgeries
- •Unable to sit and stand upright comfortably for 2 minutes
- •Unable to obtain or has contraindication to upper extremity blood pressure reading
- •Unable to perform self-measurements using the imPulse™ Tor device
- •Adults unable to consent
- •Individuals who are not yet adults (infants, children, teenagers)
- •Pregnant women
结局指标
主要结局
imPulse™ Tor System Blood Pressure Prediction
时间窗: bid/1week
Following AI/ML algorithm training and validation the difference between predicted systolic, diastolic, and mean arterial pressure predictions of the imPulse™ Tor and gold-standard measurement of an electronic blood pressure cuff will be less than 2.5 mmHg in at least 90% of the measurements, less than 5 mmHg in at least 92.5% of the measurements and less than 10 mmHg in at least 95% of the measurements.
次要结局
- imPulse™ Tor Heart Rate Prediction(bid/1week)
- imPulse™ Tor Oxygen Saturation Prediction(bid/1week)