A Clinical Study to Validate the Accuracy of BTL-001-2 ABPM (Ambulatory Non-Invasive Blood Pressure Sphygmomanometer) in Conformance with ISO 81060 – 2: 2018 + A1:2020 and A2:2024
概览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- BTL India Pvt Ltd
- 入组人数
- 200
- 试验地点
- 1
- 主要终点
- The primary objective of this study is to provide clinical
概览
简要总结
The BTL-001-2 ABPM blood pressure monitoring device is planned to be evaluated in a minimum of 200 subjects throughout all study phases that will approximate the distribution of blood pressures as outlined in ISO 81060-2:2018 + A1: 2020 and A2:2024. Two trained observers who are medical professionals will observe (listen to) the Korotkoff sounds at the brachial artery of the arm. The reference blood pressure measurements by the medical professionals will be performed sequentially with the device under test. The medical professionals will complete initial baseline auscultatory reference blood pressure measurements and device under test (BTL-001-2 ABPM) measurements. This is followed by minimum 3 and a maximum of 8 valid paired blood pressure readings of auscultatory reference blood pressure measurements alternated with device under test measurements. Once the BTL-001-2 ABPM (Ambulatory Non-Invasive Blood Pressure Sphygmomanometer) passes the clinical investigation as per the requirements ISO 81060 – 2: 2018 + A1:2020 and A2:2024 given in sections 5.1 and 5.2 (Phase 1) then it shall undergo investigation in at least 35 adult subjects according to 5.2.6 (Phase 2, ambulatory monitoring) and 5.2 in at least an additional 45 adult pregnant women (Phase 3) and 35 children of age 3-12 years (Phase 4). The prerequisite to initiate Phases 2-4 is a successful result of the Phase 1. It is expected that the data collection will take up to approximately 1 hour per subject. There is no additional follow-up required for the study. For accuracy claims, the test device measurements will be compared to those made by the reference device. The data for the final analysis will contain no fewer than 85 subjects with a minimum of 255 valid paired observations (Phase 1). Further no fewer than 35 subjects with a minimum of 11 subjects with a BP >140 mmHg will be analyzed with a minimum of 3 valid pairs in Phase 2. Also, no fewer than 45 pregnant subjects who are normotensive, hypertensive and pre-eclampsia with a minimum of 3 valid pairs will be analyzed in Phase 3. Lastly, no fewer than 35 children of age 3-12 years minimum of 3 valid pairs will be analyzed in Phase 4. At least 90% of the subjects will contribute 3 valid paired observations.
研究设计
- 研究类型
- Observational
入排标准
- 年龄范围
- 3.00 Year(s) 至 80.00 Year(s)(—)
- 性别
- All
入选标准
- •Participants grater than or equal 3 years of age
- •Arm circumference between 14 cm.
- •Willing to participate must be able to provide an informed consent or have legally authorized representative consent to participate.
- •Subjects that can tolerate a waiting period for up to 1 hour.
排除标准
- •Arm circumference less than 14 cm.
- •or greater than 50 cm.
- •Irregular heart rhythm (bigeminy, trigeminy, isolated ventricular premature beat (VPB), atrial fibrillation)
- •Korotkoffs sound K1 or K5 not audible.
- •Musculoskeletal disorder that prevents a non-invasive device to be inflated/deflated on the arm.
- •Unwilling to volunteer to participate and to sign the study specific informed consent form.
- •Patients suffering with Parkinson’s disease or other diseases with tremors.
- •Patients with coagulation disturbances.
- •Patients with serious mobility or other impairments without supervision.
- •Patients requiring urgency/emergency cardiac care
结局指标
主要结局
The primary objective of this study is to provide clinical
时间窗: 1st reading | 2nd reading | 3rd reading | 4th reading | 5th reading | 6th reading | 7th reading | 8th Reading
validation to ISO 81060-2:2018 + A1:2020 and A2:2024 for
时间窗: 1st reading | 2nd reading | 3rd reading | 4th reading | 5th reading | 6th reading | 7th reading | 8th Reading
non-invasive blood pressure measurement accuracy for the
时间窗: 1st reading | 2nd reading | 3rd reading | 4th reading | 5th reading | 6th reading | 7th reading | 8th Reading
BTL-001-2 ABPM non-invasive blood pressure monitoring
时间窗: 1st reading | 2nd reading | 3rd reading | 4th reading | 5th reading | 6th reading | 7th reading | 8th Reading
device on the intended population including Ambulatory BP
时间窗: 1st reading | 2nd reading | 3rd reading | 4th reading | 5th reading | 6th reading | 7th reading | 8th Reading
monitoring and pregnant women
时间窗: 1st reading | 2nd reading | 3rd reading | 4th reading | 5th reading | 6th reading | 7th reading | 8th Reading
次要结局
- Documenting adverse events (AEs), serious adverse events((SAEs), and device deficiencies)
研究者
Dr Meghana Murthy
Vagus Super Specialty Hospital