临床试验/NCT04737954
NCT04737954已完成不适用

A Performance Evaluation of the LumiraDx Point Of Care D-dimer Assay

LumiraDx UK Limited14 个研究点 分布在 2 个国家目标入组 679 人开始时间: 2021年3月18日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
679
试验地点
14
主要终点
Determine the accuracy of the LumiraDx POC D-Dimer assay when compared to a reference method in patients with suspected VTE.

研究概览

简要总结

In patients with symptoms of Venous Thromboembolism (VTE) (mainly deep vein thrombosis (DVT) or pulmonary embolism (PE)), the study will evaluate the agreement between the D-Dimer measurements from the LumiraDx POC D-Dimer assay and a reference method, as an aid in diagnosis, assessment and exclusion of DVT and PE.

详细描述

This is a performance evaluation study designed to assess the precision and accuracy of the LumiraDx Point of Care (POC) D-Dimer test when used in patients presenting with symptoms of Venous thromboembolism (VTE), which mainly comprises deep vein thrombosis (DVT) or pulmonary embolism (PE).

The LumiraDx POC D-Dimer test is a quantitative immunoassay providing results in under 10 minutes. The accuracy of the LumiraDx POC D-Dimer test will be assessed using capillary whole blood, venous blood, and plasma samples, by comparison to the D-Dimer results obtained from the same individuals as analysed by trained laboratory professionals using a reference device. The results of this study are intended to be used for regulatory filings for use as an in vitro diagnostic test in the European Union (EU) and other relevant geographies.

Adult males and females presenting to the study sites with symptoms of VTE will be included in the study. Approximately 1000 subjects will be recruited in total. An initial analysis will be completed on the first 200 patients to determine accuracy of the LumiraDx D-Dimer test versus the reference method. The study will continue until 1000 patients are recruited and with a minimum of 120 confirmed VTE events, in order to complete retrospective analysis using the LumiraDx D-Dimer cut-off for exclusion of VTE.

Once consent is obtained, blood tube samples will be drawn and finger-stick samples of capillary blood will be taken from the subject, these will be applied directly to unique test strips for immediate measurement of D-Dimer on the LumiraDx POC D-Dimer test.

The D-Dimer results obtained via the LumiraDx D-Dimer test will be used in a set of comparative analyses designed to determine the performance characteristics of the test. This study is an observational, cross-sectional design. As an observational study of an investigative device, results obtained during the course of the study will not inform or alter patient management or treatment decisions in any way.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Subject is greater or equal to 18 years of age Willing and able to provide written informed consent and comply with study procedures Symptoms of thromboembolic event

排除标准

  • Suspected Sinus Venous Thromboembolism / Cerebral Venous Sinus Thromboembolism End-stage renal failure on haemodialysis Current anticoagulant therapy (Fragmin, LMWH) within the last 30 days Patient has previously participated in this study Life expectancy documented as <30 days Haemodynamically unstable (e.g. cardiogenic shock) Patients taking anticoagulant therapy (DOACS, Warfarin, Heparins etc) within the last 30 days Patient deemed medically unfit to participate

研究组 & 干预措施

Suspected VTE patients

Experimental

Venous blood draw of up to 20ML and up to 6 fingerstick capillary draws

干预措施: Venepuncture (Diagnostic Test)

Suspected VTE patients

Experimental

Venous blood draw of up to 20ML and up to 6 fingerstick capillary draws

干预措施: Fingerstick (Diagnostic Test)

结局指标

主要结局

Determine the accuracy of the LumiraDx POC D-Dimer assay when compared to a reference method in patients with suspected VTE.

时间窗: 2 months

Measurement of blood samples from patients with suspected VTE in a reference method and in the LumiraDx method to assess accuracy of the LumiraDx method.

Assessment of accuracy of using the D-Dimer cut-off set by the LumiraDx D-Dimer test in excluding patients with symptoms of VTE (DVT and PE) when used in combination with the pre-probability test (Wells Score).

时间窗: 10 months

Measurement of blood samples from patients with suspected VTE in the LumiraDx D-dimer assay in conjunction with pre-test probability score and final clinical outcome in order to set a clinical cut-off for the LumiraDx D-dimer assay

次要结局

未报告次要终点

研究者

发起方
LumiraDx UK Limited
申办方类型
Industry
责任方
Sponsor