简要总结
In this study, the concentration of hemoglobin will be measured using a noninvasive pulse oximeter sensor and compared with the hemoglobin value from a blood sample of the volunteer. The accuracy of a noninvasive hemoglobin sensor(s) will be assessed by comparison to hemoglobin measurements from a laboratory analyzer.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Other
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •Greater than 18 years old
- •Exclusion criteria:
- •Do not understand the study and the risks involved
- •Deemed ineligible by study test personnel
排除标准
- 未提供
结局指标
主要结局
Accuracy of SpHb on R1-25 by Arms Calculation
时间窗: 1-5 hours per subject
Accuracy will be determined by comparing the noninvasive hemoglobin measurement of the pulse oximeter to the hemoglobin value obtained from a blood sample and calculating the arithmetic root mean square (Arms) error value. In order to obtain the Arms value, the blood sample hemoglobin value is subtracted from the pulse oximeter hemoglobin value for a number of samples, the average of this difference is computed as the bias. The standard deviation of the differences is computed as the precision. The square root of the sum of the squares of bias and precision is computed as the Arms Error value.