简要总结
The custom made device was produced by Med-El Company. It has a "cork" like stopper instead of the usual silicon ring to prevent gusher. There are two types of electrodes consisting of different length. Standard one is 25 mm (contact space 1.7 mm) and the short one is 20 mm (contact space 1.3 mm). It was used in 50 patients with different inner ear malformations. Thirteen patients had gusher, and 11 patients oozing during cochleostomy. One patient with initial prototype of the cork electrode had to be revised because of persistent oozing around the electrode. Another patient had slow extrusion of the electrode most probably due to CSF (cerebrospinal fluid) pulsation and had to be revised. Both patients had no more CSF fistula.
CSF fistula in inner ear malformations is a serious situation which may lead to recurrent meningitis. The new cochlear implant electrode with "cork" stopper looks promising in preventing the postoperative CSF leak around the electrode.
详细描述
Gusher in inner ear malformations is common in patients with incomplete partition type-I, and III. It is also common in less severe form as oozing in incomplete partition type II and large vestibular aqueduct. It is important to prevent CSF escape around the cochlear implant electrode to prevent meningitis.
The custom made device was produced by Med-El Company. It has a "cork" like stopper instead of the usual silicon ring to prevent gusher. There are two types of electrodes consisting of different length. Standard one is 25 mm (contact space 1.7 mm) and the short one is 20 mm (contact space 1.3 mm). It was used in 50 patients with different inner ear malformations.
Thirteen patients had gusher, and 11 patients oozing during cochleostomy. One patient with initial prototype of the cork electrode had to be revised because of persistent oozing around the electrode. Another patient had slow extrusion of the electrode most probably due to CSF pulsation and had to be revised. Both patients had no more CSF fistula.
CSF fistula in inner ear malformations is a serious situation which may lead to recurrent meningitis. The new cochlear implant electrode with "cork" stopper looks promising in preventing the postoperative CSF leak around the electrode.
研究设计
- 研究类型
- Interventional
- 分配方式
- Non Randomized
- 干预模型
- Single Group
- 主要目的
- Treatment
- 盲法
- None
入排标准
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Must have an inner ear malformation
- •Must meet the criteria for cochlear implantation
排除标准
- •Prior cochlear implantation surgery
结局指标
主要结局
CSF leak
时间窗: Up to 1 year
Occurrence of CSF leak after the surgery.
次要结局
- Meningitis(Up to 1 year)
研究者
Münir Demir Bajin
MD