跳至主要内容
临床试验/NCT06667375
NCT06667375已完成不适用

Description of the Performance of the "Masi" Mechanical Ventilator in a Hospital in Lima: Case Series

Fanny Lys Casado Pena1 个研究点 分布在 1 个国家目标入组 87 人开始时间: 2021年1月19日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
87
试验地点
1
主要终点
ICU discharge status

研究概览

简要总结

This study evaluates the extubation success and compares the survival rate in patients with signs of Covid-19 intubated with Masi, a novel type of mechanical ventilator, or with other conventional ventilators admitted in the ICU of a reference hospital in Lima, Peru between January and August 2021. The Masi mechanical ventilator works by automating a resuscitation bag with promising preclinical results. The key features of Masi are its low manufacturing cost, low dependence on a supply of high volumes of oxygen, low oxygen consumption, flexibility between non-invasive and invasive ventilation, and extended range of working environmental altitudes compared to commercially available ventilators. A retrospective study of variables and outcomes was done. The main goal of this study is to compare the survival rate of respiratory failure patients supported either with Masi or conventional ventilators at Hospital Vitarte II in Lima between January and August 2021.

详细描述

The translation of novel technologies for Intensive Care Units (ICU) including mechanical ventilators has been mostly about incremental improvements through increasing features assuming the presence of highly trained personnel and dedicated infrastructure such as permanent power supply and high oxygen flow. Despite the long history of use of mechanical ventilators, the Covid-19 pandemic demanded innovation to meet worldwide demands for this type of life support equipment. The inadequacy of the commercially available ventilators has been mostly attributed to weak fabrication and supply chains and inconsistencies between the required sanitary infrastructure and human resources available for which most commercial ventilators have been designed. Diverse approaches addressed the design and development of low-cost and more available mechanical ventilators worldwide achieving various levels of clinical implementation.

In Peru, the Masi (companion in Quechua) ventilator was developed based on concepts of the MIT Emergency Ventilator (MIT E-Vent) by automating a resuscitation bag. The Masi initiative showed promising pre-clinical results that allowed the production of more than 250 units under an exceptional permit to be used clinically in Peru only when no other ventilators were available. The key features of this technology are its low manufacturing cost, low oxygen consumption, flexibility between non-invasive and invasive ventilation, and extended range of working environmental altitudes compared to commercially available ventilators. Clinical characteristics, blood chemistry information, respiratory parameters, and survival rates were analyzed for 77 patients (42 ventilated by Masi, 35 ventilated by commercial ventilators).

This is an observational retrospective cohort study. The study protocol was approved by the Research Ethics Committee for Life Sciences and Technology of the PUCP (002-2021-CEICVyTech/PUCP) on September 23, 2021. A total of one hundred and three (103) patients were admitted to the ICU between January 19, 2021 until August 27, 2021, which coincides with the second Peruvian wave of the Covid-19 pandemic in the Hospital Vitarte II in Lima, Peru. Data from patients who were diagnosed with Covid-19, with signs of Acute Respiratory Distress Syndrome (ARDS), intubated, and connected to a mechanical ventilator in the ICU were de-identified for this study. From the 132 variables collected into the Electronic Health Records, only 120 selected variables were transcribed into an electronic database (Microsoft Excel V). The data was standardized and entered by a physician trained in Good Clinical Practice (Data Entry Investigator) following a Data Dictionary.

After data cleanup, this study retrospectively analyzed data from eighty-seven (87) patients who had a positive Covid-19 diagnosis and used a mechanical ventilator machine. According to the Peruvian regulatory agency, DIGEMID, Masi was indicated only when no conventional ventilator was available, and a patient needed one as determined by the ICU physician. The physicians used their criteria to prescribe Masi or other available ventilator brands.

During the described period, the ICU of the Hospital Vitarte II had seven beds and four conventional ventilators: Evolution (EVL100001-T-NF), Vela (16532-07), Newport (e360) and Dragger (Savina 300). Eight new individually validated ventilators known as Masi were provided to the ICU, and the medical personnel were trained regarding its clinical use. This device was approved to be used in Peru during the sanitary emergency by the local competent authority: Dirección General de Medicamentos, Insumos y Dispositivos Médicos (DIGEMID, Spanish acronym) based on preclinical evidence of safety.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Retrospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Patients hospitalized to ICU requiring mechanical ventilation during COVID-19 pandemic
  • Over 18 years old

排除标准

  • 未提供

结局指标

主要结局

ICU discharge status

时间窗: Patients were followed 15 days (median)

After being in the ventilator in the ICU what happened at extubation or censoring time.

次要结局

  • Ventilator parameter: Oxygen Saturation (%)(Patients were followed up 15 days (median))
  • Ventilator parameter: Respiratory rate (breaths/minute)(Patients were followed up 15 days (median))
  • Ventilator parameter: PaO2/FiO2 ratio(Patients were followed up 15 days (median))
  • Ventilator parameters: Tidal volumen (mL)(Patients were followed up 15 days (median))
  • Ventilator parameters: PIP (mmHg)(Patients were followed up 15 days (median))

研究者

发起方
Fanny Lys Casado Pena
申办方类型
Other
责任方
Sponsor Investigator
主要研究者

Fanny Lys Casado Pena

Full Professor

Pontificia Universidad Catolica del Peru

研究点 (1)

Loading locations...

相似试验