临床试验/NCT05806554
NCT05806554已完成不适用

Accuracy and Precision of the True Vie I3 Continuous Glucose Monitoring System: an Open Label, Multi-Center Trial

Sinocare6 个研究点 分布在 2 个国家目标入组 159 人开始时间: 2023年7月19日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
Sinocare
入组人数
159
试验地点
6
主要终点
Accuracy

研究概览

简要总结

The purpose of the investigation is to evaluate the performance of the I3 CGM according to the FDA's special controls for iCGM.

详细描述

Participants will attend a screening visit, 4 clinic visits with Frequent Sampling Testing (FST) at the beginning, 2 in the middle, and end of the I3 CGM sensor wear, and a visit for removal of the I3 CGM. Participants will wear 3 I3 CGM devices for up to 15 days and be instructed to continue their normal daily activities and manage their glucose as usual between in clinic FST per routine care. Participants will not have access to the I3 CGM data for the management of their blood glucose between clinic visits and will continue their usual care including CGM, if they are CGM users.

At the clinic FST visits venous blood samples will be drawn from an intravenous cannula for FST for up to 10 hours for plasma glucose determination. During the clinic visits the investigative site staff will determine the participants' insulin doses and meals and may induce both hypoglycemia and hyperglycemia of up to an hour duration.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
18 Years 至 80 Years(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • 1. Patients with T1DM or T2DM diagnosed for at least 6 months.
  • Individual is 18-80 years old, inclusive, at the time of screening.
  • Must be and have been in stable treatment regimen for at least 1 month with a multiple daily insulin dosing regimen (at least two insulin doses with meals daily) or CSII using at least two bolus doses a day with meals, irrespective of delivery device(s).

排除标准

  • 1. History of skin adhesive tolerance issues in the area of sensor placement.
  • HbA1c > 9%.
  • Insulin meal dosing based on fixed dose regimens.
  • Absence of established correction factor for high glucose.
  • Hematocrit below 10% under the lower limit of the normal range.
  • Body mass index < 18.5 kg/m
  • Inadequate intravenous access on arms.
  • Participant has had a hypoglycemic seizure within the past 6 months prior to enrollment.
  • 9. Participant has had an episode of diabetic ketoacidosis (DKA) within the past 6 months prior to enrollment.
  • 10. Participant has a history of a seizure disorder.
  • Pregnancy, planned pregnancy within the study period, or unwillingness to use reliable contraception during the study period.
  • 12. Planned MRI, CT scan or diathermia procedure for the duration of the study.

结局指标

主要结局

Accuracy

时间窗: 15 days

percentage of iCGM values in the measuring range less than 70 mg/dL (3.9 mmol/l) within +/- 15 mg/dL (0.83 mmol/l) of the

次要结局

未报告次要终点