COmparison of Clarus and Optos Ultrawide Field Imaging Systems for Geographic Atrophy (COCO-GA) Protocol

University of Wisconsin, Madison2 个研究点 分布在 1 个国家目标入组 62 人开始时间: 2023年10月6日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
入组人数
62
试验地点
2
主要终点
Comparison of geographic atrophy area measurements by Optos with standard Spectralis imaging

研究概览

简要总结

The goal of this clinical trial is to compare measurements of geographic atrophy (GA) area between several types of imaging, in order to assess accuracy. The main question to answer is which imaging device provides measurements that are most similar to the standard of care device.

Participants will be patients of a retina doctor at University Station Eye Clinic with geographic atrophy, and can expect to be in the study for 60-75 minutes.

详细描述

This study is a single site pilot comparative imaging study in which enrolled participants will undergo dilated fundus exam as per standard of care. Study images will be obtained using Optos ultrawide field fundus autofluorescence (FAF) and both blue and green ultrawide field FAF with Zeiss Clarus. Autofluorescence (AF) imaging will also be performed on the Spectralis as it is currently the gold standard. The goal will be to obtain all images in 60-75 minutes over one visit. Images will be assessed, and area of geographic atrophy measured independently by two experienced readers.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Other
盲法
None

入排标准

年龄范围
50 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • 50 years or older and may be either male or female and may be of any race
  • Established diagnosis of GA due to AMD
  • GA characteristics: GA area of between 1.25 mm² and 23 mm², with seventy percent of eyes having GA area ranging from 2.5 mm² to 17.5 mm². GA may be unifocal or multifocal. GA may be subfoveal or extrafoveal, with twenty-five percent of eyes having subfoveal GA. The presence of concurrent peripapillary atrophy will not exclude subjects from participation
  • Willing to participate as evidenced by signing the written informed consent

排除标准

  • Unable to tolerate opthalmic imaging
  • Presence of neovascular AMD on OCT as confirmed by an ophthalmologist
  • Presence of significant media opacity preventing adequate retinal imaging
  • Presence of concurrent retinal disease which may confound assessment

研究组 & 干预措施

Fundus autofluorescence (FAF) imaging

Experimental

干预措施: Optos imaging (Device)

Fundus autofluorescence (FAF) imaging

Experimental

干预措施: Spectralis FAF imaging (Device)

Fundus autofluorescence (FAF) imaging

Experimental

干预措施: Zeiss Clarus 700 imaging (Device)

结局指标

主要结局

Comparison of geographic atrophy area measurements by Optos with standard Spectralis imaging

时间窗: Up to 75 minutes

Graders will independently measure area of GA from images of each device. The measurements (in mm²) will be compared between the two devices.

Comparison of geographic atrophy area measurements between Clarus and Optos

时间窗: Up to 75 minutes

Graders will independently measure area of GA from images of each device. The measurements (in mm²) will be compared between the two devices.

Comparison of geographic atrophy area measurements by Clarus with standard Spectralis imaging

时间窗: Up to 75 minutes

Graders will independently measure area of GA from images of each device. The measurements (in mm²) will be compared between the two devices.

Comparison of Geographic Atrophy Area Measurements With Heidelberg Spectralis to Zeiss Clarus 700 and Optos California

时间窗: Up to 75 minutes

Graders will independently measure area of GA from images of each device. The measurements (in mm²) will be compared between the devices.

Comparison of Geographic Atrophy Area Measurements With Heidelberg Spectralis to Zeiss Clarus 700 and Optos California (Full Range Reported)

时间窗: Up to 75 minutes

Graders will independently measure area of GA from images of each device. The measurements (in mm²) will be compared between the devices.

Number of Eyes With Greater Than 10% Difference in Geographic Atrophy Area Than Heidelberg RegionFinder

时间窗: up to 75 mintues

次要结局

  • Prevalence of atrophy outside the standard 30 degrees in Clarus and Optos ultrawide field images(Up to 75 minutes)
  • Prevalence of other autofluorescence abnormalities outside the standard 30 degrees in Clarus and Optos ultrawide field images(Up to 75 minutes)
  • Prevalence of Atrophy Outside the Standard 30 Degrees in Clarus and Optos Ultrawide Field Images(Up to 75 minutes)
  • Prevalence of Other Autofluorescence Abnormalities Outside the Standard 30 Degrees in Clarus and Optos Ultrawide Field Images(Up to 75 minutes)

研究者

申办方类型
Other
责任方
Sponsor

研究点 (2)

Loading locations...

相似试验