Rainbow R2-25 vs. R1-25 SpHb Sub-Range Performance Equivalence

Masimo Corporation2 个研究点 分布在 1 个国家目标入组 21 人开始时间: 2009年6月18日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
21
试验地点
2
主要终点
Sub-Range Performance Equivalence of R2-25 and R1-25 Sensors by ARMS Calculation

研究概览

简要总结

In this study, the concentration of hemoglobin was measured using noninvasive pulse oximeter sensors and compared against the hemoglobin measurement from the subject's blood sample under controlled conditions. Data collected in the study was in the sub-range of 12-17 g/dL in the device's 8-17 g/dL specification range.

The purpose of the study was to evaluate a change in the sensor component and to assess performance equivalence in the sub-range. The performance of non-invasive hemoglobin is not uniform across the accuracy range.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Other

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Greater than 18 years old
  • Exclusion criteria:
  • Do not understand the study and the risks involved
  • Deemed ineligible by study test personnel

排除标准

  • 未提供

结局指标

主要结局

Sub-Range Performance Equivalence of R2-25 and R1-25 Sensors by ARMS Calculation

时间窗: 1-5 hours

Sub-range performance equivalence was determined by comparing the noninvasive hemoglobin measurement of the pulse oximeter sensor to the hemoglobin value obtained from a blood sample and calculating the arithmetic root mean square (Arms) error value. In order to obtain the Arms value, the blood sample hemoglobin value is subtracted from the pulse oximeter hemoglobin value for a number of samples, the average of this difference is computed as the bias. The standard deviation of these differences is computed as the precision. The square root of the sum of the squares of bias and precision is computed as the Arms Error value.

次要结局

未报告次要终点

研究者

申办方类型
Industry
责任方
Sponsor