Results of the SpO2 and Pulse Rate Accuracy Comparison of VitalDetect™ to Arterial Blood CO-Oximetry and Reference ECG

Vital USA, Inc.2 个研究点 分布在 1 个国家目标入组 10 人开始时间: 2019年11月4日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
10
试验地点
2
主要终点
Evaluate SpO2 Accuracy Per Standards for Pulse Oximetry.

研究概览

简要总结

Clinical Investigation Plan: Vital USA SpO2 Accuracy Comparison to Arterial Blood CO-Oximetry Study ID# PR 2019-353, Version: Rev 1, Date: 15 Oct 2019 Study Dates: November 4-5, 2019

详细描述

An SpO2 accuracy comparison was conducted as part of the final validation of the VitalDetect™ pulse oximetry system. The study was conducted in accordance to Code of Federal Regulations (CFR) for Non-Significant Risk (NSR) investigational studies, following International Standards Organization (ISO) 14155:2011 as appropriate and the pulse oximetry guidelines of ISO 80601-2-61:2017 applicable sections, and Pulse Oximeters - Premarket Notifications Submissions [510(k)s] Guidance For Industry and Food and Drug Administration Staff (issued: March 4, 2013).

The purpose of this study was to validate the SpO2 accuracy of the VitalDetect™ pulse oximetry system during non-motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry. It was expected that the Accuracy Root Mean Square (ARMS) performance of the oximetry system would meet the required specification of ARMS of 3.0% or less.

A secondary goal was to evaluate the pulse rate performance simultaneously collected over the SpO2 range covered. The Pulse Rate Accuracy Root Mean Square (Arms) performance of the VitalDetect™ pulse oximetry system, were expected to meet a specification of 3 bpm.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Device Feasibility
盲法
None

入排标准

年龄范围
18 Years 至 52 Years(Adult)
性别
All
接受健康志愿者

入选标准

  • 未提供

排除标准

  • 未提供

结局指标

主要结局

Evaluate SpO2 Accuracy Per Standards for Pulse Oximetry.

时间窗: 90 Seconds

This study was a comparative, single-center, non-randomized study conducted to evaluate the SpO2 accuracy per standards and guidelines identified above for SpO2 accuracy for pulse oximetry equipment over the range of 70-100% SaO2 under non-motion conditions. Arterial blood sampling measured by functional SaO2 CO-Oximetry, was used as the basis for comparison. As a secondary goal, the pulse rate accuracy was compared to ECG Heart Rate realized over the range throughout the hypoxia procedure. Testing was conducted under normal office environment conditions.

次要结局

  • The Pulse Rate Accuracy Compared to ECG Heart Rate.(90 Seconds)