SpO2 Validation of Noninvasive Red Diamond Disposable Pulse Oximeter Sensor
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 35
- 试验地点
- 2
- 主要终点
- Accuracy of Red Diamond Disposable Pulse Oximeter Sensor by Arms Calculation
研究概览
简要总结
In this study, the level of oxygen within the blood will be reduced in a controlled manner by reducing the concentration of oxygen the study volunteer breathes. The accuracy of a noninvasive pulse oximeter sensor will be assessed by comparison to the oxygen saturation measurements from a laboratory blood gas analyzer.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Other
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 50 Years(Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- 未提供
排除标准
- 未提供
研究组 & 干预措施
Test Subject
All subjects are enrolled into the test group and all subjects receive the noninvasive Red Diamond Disposable Pulse Oximeter Sensor
干预措施: Red Diamond Disposable Pulse Oximeter Sensor (Device)
结局指标
主要结局
Accuracy of Red Diamond Disposable Pulse Oximeter Sensor by Arms Calculation
时间窗: 1-5 hours per subject
Accuracy will be determined by comparing the noninvasive blood oxygen saturation measurement of the pulse oximeter to that obtained from a blood sample and calculating the Accuracy root mean square (ARMS) error value. In order to obtain the Arms value, the blood oxygen saturation measurement form a laboratory pulse Co-Oximeter is subtracted from the pulse oximeter oxygen saturation measurement for a number of samples, the average of this difference is computed as the bias. The standard deviation of the differences is computed as the precision. The square root of the sum of the squares of bias and precision is computed as the ARMS Error value.
次要结局
未报告次要终点