BiOSENCY Bora Band Respiration Rate Development and Validation Study
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- Biosency
- 入组人数
- 27
- 试验地点
- 2
- 主要终点
- Accuracy of BORA Band respiratory rate measurements versus reference end-tidal CO2
研究概览
简要总结
The purpose of this study is to conduct a Respiratory Rate accuracy validation comparing the BORA BAND™ Wristband Monitoring System to the Reference. Respiratory Rate will be compared to an FDA cleared End Tidal Carbon Dioxide monitor (GE Datex-Ohmeda). The testing will be conducted in two phases. The first phase will be an initial data collection to assist in the algorithm development of respiration rate. Upon a successful Phase one performance, the second phase will be the Respiratory Rate Validation conducted at a different period in time.
详细描述
For phase 1, there will be 5 to 6 volunteer test subjects. For the final validation phase, there will be a minimum of 20 volunteer test subjects. All subjects will be 18 or older and selected to represent a range of body types including small, average, muscular, and large with a range of BMI's.
Each subject will be connected to a commonly used, FDA cleared ECG Impedance monitor (to be used for informational purposes) and an End Tidal Carbon Dioxide monitor (GE Datex-Ohmeda) and BiOSENCY BORA BAND™ wristband monitoring system. The end tidal CO2 monitor will be attached to the patient to determine performance of respiratory rate metrics (Reference). Each subject will be instrumented with a mouthpiece that allows for measurement of the end tidal carbon dioxide (EtCO2) respiratory rate (Reference) and tidal volume.
A range of stable respiratory rates will be elicited from each volunteer test subject. The rates will be approximately 5, 10, 15, 20, 25, 30, 35, 40, 45 and 50 breaths per minute; with some natural variation from these exact numbers. Paced breathing app will be used. Not all subjects will be able to breathe at a constant level for every rate.
The respiratory rate will be measured simultaneously with EtCO2 (Reference) and the Devices Under Test.
For the data analysis, the EtCO2 monitor will be used to assess the stability of the data. If the respiratory rate varies by more than +3 bpm during a one-minute period of the target respiration rates, then the data for that period will be considered unstable and removed from analysis.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Diagnostic
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •Subject must have the ability to understand and provide written informed consent
- •Subject is adult over 18 years of age
- •Subject must be willing and able to comply with study procedures and duration
- •Subject is a non-smoker
- •Male or female of any race
- •Wrist circumference with the range of 5.5-8 inches (13 to 21cm)
排除标准
- •Compromised circulation, injury, or physical malformation of fingers, toes, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. Tattoo in the optical path which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the particular sites utilized.)
- •Subjects with known respiratory conditions such as:
- •uncontrolled / severe asthma,
- •pneumonia / bronchitis,
- •shortness of breath / respiratory distress,
- •respiratory or lung surgery,
- •emphysema, COPD, lung disease
- •Subjects with self-reported heart or cardiovascular conditions such as:
- •have had cardiovascular surgery
- •Chest pain (angina)
- •heart rhythms other than a normal sinus rhythm or with respiratory sinus arrhythmia
- •previous heart attack
- •blocked artery
- •unexplained shortness of breath
- •congestive heart failure (CHF)
- •history of stroke
- •transient ischemic attack
- •carotid artery disease
- •myocardial ischemia
- •myocardial infarction
- •cardiomyopathy
- •Self-reported health conditions as identified in the Health Assessment Form
- •diabetes,
- •uncontrolled thyroid disease,
- •kidney disease / chronic renal impairment,
- •history of seizures (except childhood febrile seizures),
- •epilepsy,
- •history of unexplained syncope,
- •recent history of frequent migraine headaches,
- •recent head injury within the last 2 months,
- •Cancer / chemotherapy
- •Other known health condition, should be considered upon disclosure in health assessment form
研究组 & 干预措施
Respiratory rate accuracy measurement
干预措施: Measurement of respiratory rate with DUT and comparison with reference (Device)
结局指标
主要结局
Accuracy of BORA Band respiratory rate measurements versus reference end-tidal CO2
时间窗: 5 days
Accuracy of the BORA Band respiratory rate measurement as compared to reference end-tidal CO2. Results shall include: * Respiratory rate root mean square accuracy of BORA Band * Respiratory rate bias of BORA Band
次要结局
未报告次要终点