Clinical Study About Fat Measurement on Multifunction Keito by Comparison Versus Its Predicate K014009

Aguiflai Iberica, S.L.1 个研究点 分布在 1 个国家目标入组 177 人开始时间: 2007年6月最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
177
试验地点
1

研究概览

简要总结

The researchers believe that the multifunction Keito devices shares with TANITA TBF-300 (K014009) the same efficacy on the body fat measurement. Even if the body contact points are not the same, the researchers deeply believe that the tetrapolar BIA used by the predicate gives results as similar as the multifunction keito does.

详细描述

177 volunteers were enrolled in this clinical Study. Before taking the fat measurement by the devices under test, the observers measured the weight and height of each volunteer prior to measure the body fat. This was done to configure a preliminary idea about the body composition of the subject.

Both devices are designed by the same technology to emit a low current through the body, and finally make a measurement of the impedance found between the body contact points. We expected the results to be very similar.

研究设计

研究类型
Observational
观察模型
Ecologic Or Community
时间视角
Cross Sectional

入排标准

年龄范围
18 Years 至 82 Years(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • All population who meet with the requirements described on the user's manual.

排除标准

  • Pregnant women.
  • Individuals who have a pacemaker or any other internal medical device.
  • Population whose weight is under 15 Kg and less than 135 cm tall.

研究者

发起方
Aguiflai Iberica, S.L.