NCT01325922已完成不适用
50/50% Tilt and Tune Defibrillation Waveforms Utilizing High Voltage Lead Integrity Check Impedance
适应症
干预措施
试验速览
- 阶段
- 不适用
- 状态
- 已完成
研究概览
简要总结
Study Description The purpose of this study is to compare the defibrillation efficacy between the 50/50% tilt biphasic waveform and the Tuned biphasic waveform utilizing High Voltage Lead Integrity Check (HVLIC) impedance.
Hypothesis The Tuned waveform reduces defibrillation thresholds (DFTs) when compared to the 50/50% tilt waveform.
Study Methods
- This is an acute, paired-sample, randomized (waveform testing sequence) study
- Patients are implanted with an Food and Drug Administration (FDA) approved St. Jude Medical (SJM) EpicTM+ DR/VR Implantable Cardioverter Defibrillator (or any other standard output ICD after Epic+) and defibrillation lead system.
- All study testing is completed at implant.
- Patients are randomized to begin DFT testing with either the 50/50% tilt or Tuned waveform.
- The HVLIC feature is used to measure the high voltage lead impedance and to program the Tuned and 50/50% tilt waveform pulse widths.
- The DFT is determined by an optimized binary search method that is designed to reduce the overall number of VF inductions to an average of 5.13 while maintaining very high resolution and accuracy.
- Total # of centers - 15 centers
- Sample size - 60 patients
研究设计
- 研究类型
- Interventional
- 分配方式
- Randomized
- 干预模型
- Parallel
- 主要目的
- Other
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Patient is a candidate for ICD implantation.
- •Patient is able to tolerate DFT testing.
排除标准
- •Patient has a mechanical valve in the tricuspid position.
- •Patient is pregnant.
- •Patient is less than 18 years old.
研究组 & 干预措施
50/50% Tilt
Other
干预措施: DFT Testing with a 50/50% tilt (Procedure)
研究者
相似试验
已完成
不适用
Comparison of Fixed Tilt and Tuned Waveforms in Right-sided ImplantsHealthyNCT00873691Abbott Medical Devices129
已完成
不适用
Comparison of 50% Tilt and Tuned Waveforms in Single-Coil Active Can ConfigurationSudden Cardiac DeathNCT00874445Abbott Medical Devices77
已完成
不适用
Defibrillation of ECG rhythms - a comparison and decision studycardiac arrestDRKS00031312Klinik für Anästhesiologie und Operative Intensivmedizin der Universität zu Köln (AöR)207
已完成
不适用
Comparison of Efficacy and Safety of Two Biphasic Defibrillators in Cardioversion of Atrial FibrillationArrythmiaElectric CountershockAtrial FibrillationNCT04032678University National Heart Hospital75
终止
不适用
Acute Defibrillation StudyVentricular ArrhythmiasNCT02227121Medtronic Cardiac Rhythm and Heart Failure16