Clinical Performance of the AFGen1 Device Over a 7-day Period

TriVirum, Inc.1 个研究点 分布在 1 个国家目标入组 15 人开始时间: 2023年10月5日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
15
试验地点
1
主要终点
Adhesive performance

研究概览

简要总结

AFGen1 is indicated for use on symptomatic or asymptomatic adults who are at risk of developing or who have atrial fibrillation, where a software assisted analysis of ambulatory ECG is needed to identify episodes of Afib. The purpose of this study is to establish the clinical performance of AFGen1 on human participants.

详细描述

The primary purpose of this study is to demonstrate the ECG signal acquired by the AFGen1 device is of adequate quality and is suitable to support its intended use for the detection of Afib. This will be evaluated by comparing the ECG data acquired by AFGen1 to that acquired by an FDA cleared 12-lead ECG device. The study will also seek to demonstrate that there is no degradation in the quality of the ECG signal it acquires over the maximum 7-day wear period that it can be prescribed for use.

The secondary purpose of this study is to confirm the adhesive performance of the device is appropriate to support its intended use per the requirements of sections 4.4 and 5.4 of ANSI AAMI EC12-1 .

研究设计

研究类型
Observational
观察模型
Case Only
时间视角
Prospective

入排标准

年龄范围
18 Years 至 90 Years(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • 18+ willing to sign the consent form

排除标准

  • Implanted pacemakers
  • Implanted cardioverter defibrillators
  • Implanted cardiac resynchronization devices
  • Potential life-threatening arrythmias
  • Physical or mental health conditions that would prevent the person from being able to follow instructions regarding participation in the study
  • Open wounds, abraded or irritated skin at the application site
  • Planned to undergo a MRI during the course of the study duration
  • Known or suspected to be pregnant
  • Student or employee of TriVirum

结局指标

主要结局

Adhesive performance

时间窗: 7 days

For the adhesive performance element of the study, the overall average wear time for the device must be greater than 5-days.

Quantitative Evaluation

时间窗: 7 days

For the quantitative evaluation of the mean and standard deviation of the QRS amplitude, duration, and timing correlations for all study subjects and for the end of the wear period for those relevant subjects, the mean correlation coefficient must be greater than 0.95 and standard deviation must be less than 0.05 to validate a 95% confidence interval with significant factor of assurance. validate a 95% confidence interval with significant factor of assurance.