Thales Thermography Triage (3T) - Pilot Project: Developing a Thermal Camera That Accurately Records Participants' Temperature From Images of Their Face
试验速览
- 阶段
- 不适用
- 入组人数
- 200
- 试验地点
- 1
- 主要终点
- To determine the diagnostic accuracy of HTD for the measurement of fever, using an equivalence comparison with the maximum temperature recorded with two routinely used temperature measurements (tympanic and forehead), margin ±0.2°C.
研究概览
简要总结
COVID Related Study.
This study will compare the Thales High Temperature Detection (HTD) system with the two routinely used temperature measurement systems (tympanic and forehead) in patients and staff in the Emergency Department.
The aims of the study are:
- To determine the diagnostic accuracy of HTD for the measurement of fever, using an equivalence comparison with the maximum temperature recorded with two routinely used temperature measurements (tympanic and forehead), margin ±0.2°C.
- To determine if skin tone is an independent confounding factor in the measurement of fever using HTD
- To establish performance of internal calibration of Thales device
- To power for a definitive Clinical Investigation to determine clinical efficacy and safety cut off with HTD If results are encouraging the aim would then be to undertake a clinical investigation in order to get CE marking for use in a clinical setting.
详细描述
In the fifty years since the emergence of thermal imaging technology, Thales in Glasgow has built up a world leading capability in the design, manufacture and supply of Thermal Imaging cameras. In addition to the cameras, Thales in Glasgow has a particular expertise in developing image processing algorithms (conventional and artificial intelligence based) to allow the cameras to perform critical user tasks beyond mere imaging.
COVID-19 [SARS-COV-2] has placed a huge challenge on the world. In response to the crisis, Thales is engaged in a number of initiatives, including one aimed at the possible application of thermal imaging cameras to detect people with a fever and hence those who may be suffering from COVID-19.Fever is a key symptom of covid-19. One of the major limitations of existing systems is thermographic accuracy. Most current thermographic camera systems offer an accuracy of ± 0.5°C. This can be reduced to ±0.3°C with the inclusion of a "Black Body" calibration source, although this limits the portability of the system. Using their unique knowledge and expertise in high resolution and precision thermal imaging and thermal image processing, Thales aims to develop a fever screening system with an accuracy of ± 0.1°C to enable more effective identification and triage of people with fever.
What Thales lacks is real world thermal images of patients suffering fever and access to clinicians who can advise on the medical aspects of the work. This research would fill these two voids.
In a pilot study conducted in the Emergency Department, Queen Elizabeth University Hospital (20/NS/0064), the Thales High Temperature Detection (HTD) system was found to be more accurate than a tympanic thermometer which captures Infrared radiation emitted by the tympanic membrane in the ear or forehead thermometer. Tympanic temperature measurements are typically used in the hospital setting in Scotland whereas forehead IR measurement is common in transportation and workplace settings.
The HTD has been redesigned in response to three key findings from the pilot study:
研究设计
- 研究类型
- Observational
- 观察模型
- Cohort
- 时间视角
- Prospective
入排标准
- 年龄范围
- 16 Years 至 —(Child, Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •Patients ≥ 16 years old with no upper age limit
- •Patients able to read and understand English
- •Patients able to give informed consent
- •Patients being triaged through ED for any complaint, not necessarily COVID-19 QEUH ED staff
排除标准
- •Patients not meeting the inclusion criteria
- •Patients who do not have capacity to consent
- •Patients attending ED who are fast-tracked without triage
结局指标
主要结局
To determine the diagnostic accuracy of HTD for the measurement of fever, using an equivalence comparison with the maximum temperature recorded with two routinely used temperature measurements (tympanic and forehead), margin ±0.2°C.
时间窗: First 4 hours of hospitalisation.
Measurement of temperature in celsius
次要结局
未报告次要终点
研究者
David Lowe
Dr David J Lowe
NHS Greater Clyde and Glasgow