The Gambro Polyflux HD-C4 Big Ease Of Use And Thrombogenicity Study
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 33
- 试验地点
- 2
- 主要终点
- Ease of Use: Priming Blood Side
研究概览
简要总结
The purpose of this study is to collect information on the ease of use and thrombogenicity of the Polyflux HD-C4 Big and Fresenius Optiflux 180NR or 200NR dialyzers under conditions of routine clinical use for hemodialysis.
详细描述
This ease of use study involves the use of FDA 510(k) cleared hemodialyzers and will have no impact on the patient's routine dialysis therapy. This study is aimed at obtaining feedback from the nursing and technical staff regarding the use and thrombogenicity of the Polyflux HD-C4 Big, and Fresenius Optiflux 180NR or 200NR dialyzers. The dialysis nursing staff will complete ease of use and thrombogenicity assessments following routine dialysis therapy. Any adverse events will be tracked for the study patients.
研究设计
- 研究类型
- Observational
- 观察模型
- Other
- 时间视角
- Prospective
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Hemodialysis patients 18 years of age or older
排除标准
- 未提供
结局指标
主要结局
Ease of Use: Priming Blood Side
时间窗: Priming at each treatment
For every study treatment/dialyzer, the nursing staff rated the ease of priming blood side as follows: Very Easy:Air is easily removed without knocking or clamping procedures. Acceptable:Knocking and/or clamping required for efficient air removal. Difficult:Knocking and/or clamping and additional volume of saline or extended recirculation needed to remove air. Very Difficult:Air could not be removed.
Dialyzer Assessment: Fiber Condition
时间窗: Each treatment: assessment of the condition of the dialyzer fibers after rinse-back
For every study treatment/dialyzer, the nursing staff rated the condition of the dialyzer fibers after rinse-back as follows: Very Good:All fibers appear white. Good:Few fibers (less than 10) appear PINK / RED. (check one) Poor:Several fibers (more than 10) appear PINK / RED. (check one) Very Poor:Most fibers (more than 75%) appear PINK / RED. (check one)
Dialyzer Assessment: Arterial Header Condition
时间窗: Each treatment: assessment of the condition of the arterial header after rinse-back
For every study treatment/dialyzer, the nursing staff rated the condition of the arterial header after rinse-back as follows: Very Good:Surface as clean as expected. Good:Distinct red ring appears at the dialyzer wall. Poor:Significant appearance of clots, randomly distributed on the surface. Very Poor:Completely red and/or covered with clots.
Ease of Use: Priming Dialysate Side
时间窗: Priming at each treatment
For every study treatment/dialyzer, the nursing staff rated the ease of priming dialysate side as follows: Dialysate Side at 500 ml Priming Volume (check one): Perfect: No air visible. Acceptable: Some air visible, but considered insignificant. Not Acceptable: Additional actions needed to sufficiently remove air.
Dialyzer Assessment: Venous Header Condition
时间窗: Each treatment: the assessment of the condition of the venous header after rinse-back
For every study treatment/dialyzer, the nursing staff rated the condition of the venous header after rinse-back as follows: Very Good:Surface as clean as expected. Good:Distinct red ring appears at the dialyzer wall. Poor:Significant appearance of clots, randomly distributed on the surface. Very Poor:Completely red and/or covered with clots.