New Doppler Instrument for Preventing Impending Access-Graft Failure Randomized Trial
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 111
- 试验地点
- 4
- 主要终点
- Number of low flow events
研究概览
简要总结
This research study aims to show that a newly developed investigational Doppler instrument can reliably measure blood flow in dialysis access grafts, to see if a decrease in the flow may be a sign of impending graft failure and also to see if there is evidence that by monitoring the flow in these grafts their lives may be extended.
The new investigational device is to be known as FloMon; this is an ultrasonic device that provides blood flow measurements from a small probe placed in a noninvasive manner above the graft or blood vessel. The instrument is being developed so that it may help give insight to on-coming graft failure which may be prevented.
The FloMon is an extension of a previously Food and Drug Administration (FDA) cleared instrument, Echoflow-A, which had been used in a similar study as this one. The Echoflow-A is similar to this device in all respects with the exception of design and calculation of measurement. The FloMon compared to the Echoflow-A is better ergonomically designed and reads flow readings in about 2 minutes versus twice the time necessary to take similar measurements using the Echoflow-A.
详细描述
Vascular access failure is one of the largest causes of morbidity for chronic hemodialysis patients: 16-25% of hospital admissions among United States end-stage renal disease (ESRD) patients are related to vascular access complications, and the associated cost is estimated to be over one billion dollars per year.
The most frequent cause of AV access-graft failure is thrombosis that occurs as a result of stenosis at or near the venous anastomosis. Over 50% of existing vascular access-grafts in the US are 6 mm PTFE tubes and their thrombosis rate is much higher than that of native Arteriovenous Fistulas (AVF).
Several studies have shown that early detection and repair of stenosis can prevent access thrombosis, reducing the morbidity and costs that follow thrombosis. The cost of duplex ultrasound for access monitoring is prohibitive. Blood flow, measured by indicator-dilution methods, seems to be a reliable indicator of impending access failure, but recent randomized controlled studies intended to examine this were equivocal.
This study, funded by the NIH (2R44 DK067775), is to determine if a new instrument to monitor access blood-flow can reduce graft failure and associated costs. In a pilot study using the new Doppler ultrasound instrument for weekly monitoring, conducted by the Renal Research Institute and funded by the NIH, the method predicted impending graft failure with 80% sensitivity and a false-alarm rate of less than .5/year. It measures access blood flow in less than three minutes, and uses almost no consumables.
This study is designed to test if monitoring with the new instrument improves health and reduces cost for hemodialysis patients with access-grafts.
研究设计
- 研究类型
- Interventional
- 分配方式
- Randomized
- 干预模型
- Parallel
- 主要目的
- Prevention
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Chronic hemodialysis patients with an AV access graft
排除标准
- •Anticipated change in renal replacement modality or geographic location
- •Inability to give informed consent
- •Anticipated life expectancy of less than one year
研究组 & 干预措施
Weekly measurement of graft flow
The surveillance group was measured with the FloMon instrument weekly, and all procedures related to their access-grafts recorded
干预措施: Weekly measurement of graft flow (Other)
Control
The control group was questioned weekly with regard to their graft status and whether there had been any graft-related procedures
结局指标
主要结局
Number of low flow events
时间窗: Mean follow-up of 15 months
A "Low-Flow Event" is an intervention, whether an angioplasty or thrombectomy, to increase graft flow.
次要结局
未报告次要终点