临床试验/NCT06912516
NCT06912516已完成不适用

PULSE OXIMETRY Performance Test for Vitacam. Controlled Desaturation Test.

Vitacam2 个研究点 分布在 1 个国家目标入组 24 人开始时间: 2023年12月22日最近更新:

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
24
试验地点
2
主要终点
Accuracy of oxygen saturation measurements - Root Mean Square Error

研究概览

简要总结

This clinical investigation aims to collect data for algorithm development and testing of Vitacam as well as to later validate the system using a simultaneously collected, but segregated, hold-out validation dataset. Vitacam is a software medical device for measuring vital signs, manufactured by NE Device SW Oy. It analyses digital video recorded using a smart device with a camera to locate a subject's face and chest region. From both these regions of interest, the video is further analysed to obtain the heart rate, heartbeat intervals, respiratory rate and oxygen saturation. These measurements are used as inputs to clinical decision making. Vitacam uses remote photoplethysmography to obtain the heart rate and heartbeat intervals, and will build on this to obtain oxygen saturation using information from red-green-blue channels.

The main focus of this study is to validate the oxygen saturation, heart rate and heartbeat intervals measured by Vitacam against gold standard references, applying the ISO 80601:2019 standard.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Other
盲法
None

入排标准

年龄范围
18 Years 至 40 Years(Adult)
性别
All
接受健康志愿者

入选标准

  • Age: 18-40 years
  • ASA category 1
  • Positive Allen's test
  • Passing the physical examination
  • Ability to give informed consent
  • Volunteering for the study
  • Fluent Finnish or English

排除标准

  • Age under 18 years or over 40 years
  • Pregnancy
  • Does not pass the physical examination
  • Negative Allen's test
  • Relevant medical illnesses
  • Cardiovascular disease
  • Pulmonary disease
  • Inability to give informed consent

结局指标

主要结局

Accuracy of oxygen saturation measurements - Root Mean Square Error

时间窗: A 30-45 minute period during which the participant is subjected to the de-saturation process

Accuracy as measured by Root Mean Square Error should be less than or equal to 4% for SpO2 as per ISO 80601-2-61:2019. It is calculated from the difference of the SpO2 measured from the investigational device and that of the reference device.

Accuracy of oxygen saturation measurements - Mean absolute error

时间窗: A 30-45 minute period during which the participant is subjected to the de-saturation process

Accuracy as measured by Mean Absolute Error should be less than or equal to 4% for SpO2 as per ISO 80601-2-61:2019. It is calculated from the difference of the SpO2 measured from the investigational device and that of the reference device.

次要结局

  • Accuracy of interbeat intervals of the investigational device against successive R waves of the QRS complex from an Electrocardiogram(A 30-45 minute period during which the participant is subjected to the de-saturation process)

研究者

发起方
Vitacam
申办方类型
Industry
责任方
Sponsor