跳至主要内容
临床试验/NCT05378347
NCT05378347招募中不适用

Endurant Stent Graft System vs Excluder Endoprothesis: A Global, Prospective, Randomized Clinical Trial in Sac Regression

Medtronic Cardiovascular68 个研究点 分布在 8 个国家目标入组 600 人开始时间: 2023年1月5日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
招募中
发起方
入组人数
600
试验地点
68
主要终点
Sac Regression

研究概览

简要总结

The purpose of this trial is to generate clinical evidence related to key performance outcomes of Endurant II/IIs Stent Graft Systems verses Gore Excluder / Excluder Conformable AAA Endoprosthesis in subjects with Abdominal Aortic Aneurysms. Subjects are randomized and imaging collected at all follow-up time points to assess the primary endpoint.

详细描述

This is a post-market, prospective, interventional, global, multi-center, randomized, dual-arm clinical trial. The primary objective of this trial is to evaluate sac regression outcomes of the Medtronic Endurant II/IIs Stent Graft System and Gore Excluder/Excluder Conformable AAA Endoprosthesis in standard EVAR subjects. Subjects are randomized on a 1:1 basis to receive EVAR with either a Medtronic Endurant II/IIs Stent Graft Systems or Gore Excluder / Excluder Conformable AAA Endoprosthesis. Data is collected pre- and post-procedure at discharge, at 30 days, and once a year until the 5-year follow-up is completed. Approximately 600 subjects are planned to be enrolled at up to 100 sites globally.

Product Names:

  • Medtronic Endurant II/IIs Stent Graft System
  • Gore Excluder and/or Gore Excluder Conformable AAA Endoprosthesis Imaging is collected for all follow-up time points and are evaluated by a core lab for the key primary and secondary objectives.

研究设计

研究类型
Interventional
分配方式
Randomized
干预模型
Parallel
主要目的
Treatment
盲法
Single (Outcomes Assessor)

盲法说明

Clinical Events Committee members as well as key sponsor team members will be blinded

入排标准

年龄范围
20 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Subject and the treating physician agree that the subject will return for all required follow up visits
  • Subject or legal representative or consultee, as applicable, has consented for trial participation and signed the Informed Consent approved by the sponsor and by the Ethics Committee/Institutional Review Board
  • Subject has an aneurysm diameter of ≥ 5 cm (if woman) ≥ 5.5 cm (if man) or as stated otherwise in regional addenda.
  • Subject's AAA anatomy is appropriate for both Medtronic Endurant II/IIs Stent Graft System and Gore Excluder/Excluder Conformable AAA Endoprosthesis as per assessment of both treating physician and Core Lab in accordance with the overlapping commercially available IFUs per applicable region.

排除标准

  • Subject is participating in an investigational drug or device study which may bias or interfere with the endpoints and follow-up of this trial
  • Subject has an estimated life expectancy of ≤ 3 years as judged by the investigator
  • Subject has an aneurysm that is:
  • Suprarenal/pararenal/juxtarenal
  • Isolated ilio-femoral
  • Inflammatory
  • Pseudoaneurysm
  • Concomitant or prior dissection involving the abdominal aorta or iliac arteries
  • Symptomatic AAA
  • Subject has significant thrombus and / or calcium at the arterial implantation sites, specifically the proximal aortic neck and distal iliac artery interface. Significant thrombus may be quantified as thrombus ≥ 2 mm in thickness and / or ≥ 25% of the vessel circumference in the intended seal zone of the aortic neck.
  • Subject requires emergent aneurysm treatment, for example, trauma or rupture
  • Subject with connective tissue disease that may have caused the aneurysm e.g. Marfan syndrome, Ehlers-Danlos, Loeys-Dietz syndrome
  • Subject has previously undergone surgical or endovascular treatment in the abdominal aorta or the iliac arteries for aneurysm or occlusive disease
  • Planned use of aorto-uni-iliac (AUI) main body device
  • Any planned additional device (apart from the main body, limb stent graft and extensions per assigned treatment per randomization) during index or staged procedure, e.g., endostaple or anchor, Iliac branch endoprosthesis, embolization, etc.
  • Planned coverage of the internal iliac artery/arteries
  • Subject has an estimated glomerular filtration rate (eGFR) < 45 ml/min/1.73m2 or subject is on dialysis
  • Subject has a systemic infection who may be at increased risk of endovascular graft infection, per investigator's discretion
  • Subject has a psychiatric or other condition that may interfere with the trial, per investigator's discretion
  • Subject is of childbearing potential in whom pregnancy cannot be excluded
  • Subject has a known hypersensitivity or contraindication to anticoagulants, anti-platelets, or contrast media, which is not amenable to pre-treatment
  • Subject belongs to a vulnerable population per investigator's judgment
  • Subject has an active COVID-19 infection or relevant history of COVID- 19

研究组 & 干预措施

Medtronic Endurant II/IIs

Experimental

Subjects are randomized on a 1:1 basis to receive an EVAR procedure with either a Medtronic Endurant II/IIs endoprothesis or Gore's Excluder / Excluder Conformable.

干预措施: Medtronic Endurant II or Endurant IIs Stent Graft System (Device)

Gore Excluder / Excluder Conformable

Experimental

Subjects are randomized on a 1:1 basis to receive an EVAR procedure with either a Medtronic Endurant II/IIs endoprothesis or Gore's Excluder / Excluder Conformable.

干预措施: Gore Excluder or Gore/ Excluder Conformable AAA Endoprosthesis (Device)

结局指标

主要结局

Sac Regression

时间窗: 12 months

Proportion of patients with sac regression at 1 year, defined in accordance with SVS guidelines as the reduction in maximum diameter by \>= 5 mm when compared to the first CT imaging study obtained within 90 days after index procedure.

次要结局

  • Secondary intervention incidence rate(30 days, 12 months and annually to 5 years)
  • Type II endoleak incidence rate(30 days, 12 months and annually to 5 years)
  • Type I endoleak incidence rate(30 days,12 months and annually to 5 years)
  • All cause mortality incidence rate(30 days, 12 months and annually to 5 years)
  • Aneurysm sac change by volume incidence rate(12 months and annually to 5 years)
  • Aneurysm sac change by diameter as a continuous variable(12 months and annually to 5 years)

研究者

发起方
Medtronic Cardiovascular
申办方类型
Industry
责任方
Sponsor

研究点 (68)

Loading locations...