NCT02262195已完成不适用
Dysfunctional Hemoglobin Pulse Co-ox
适应症
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 62
- 试验地点
- 1
- 主要终点
- Evaluate carboxyhemoglobin over the range of 0-15 percent as assessed by co-oximetry
研究概览
简要总结
The purpose of the study is to determine if a new pulse oximeter accurately detects carboxyhemoglobin and methemoglobin when intentionally increased in the blood.
研究设计
- 研究类型
- Observational
- 观察模型
- Cohort
- 时间视角
- Prospective
入排标准
- 年龄范围
- 18 Years 至 45 Years(Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •Subjects does not have significant medical problems
- •Subject is willing to provide written informed consent
- •Subject is between 18 and 45 years of age
排除标准
- •Has a BMI greater than 31
- •Has had any relevant injury at the sensor location site
- •Has deformities or abnormalities that may prevent proper application of the device under test
- •Has a known respiratory condition
- •Is currently a smoker
- •Has a known heart or cardiovascular condition
- •Is currently pregnant
- •Is female and actively trying to get pregnant
- •Has a clotting disorder
- •Is known to have a hemoglobinopathy such as (anemia, bilirubinemia, sickle-cell anemia, inherited or congenital methemoglobinemia)
- •The subject has a COHb greater than 3% or MetHb greater than 2%
- •Has taken blood thinners or medication with aspirin within the last 24 hours
- •Has unacceptable collateral circulation from the ulnar artery
- •Has donated more than 300 mL of blood within one month prior to start of study
- •Is unwilling or unable to provide written informed consent to participate in the study
- •Is unwilling or unable to comply with the study procedures
- •Has another health condition which in the opinion of the principal investigator makes him/her unsuitable for testing
结局指标
主要结局
Evaluate carboxyhemoglobin over the range of 0-15 percent as assessed by co-oximetry
时间窗: 4 hours
Evaluate methemoglobin over the range of 0-15 percent as assessed by co-oximetry
时间窗: 4 hours
次要结局
- Determine oxygen saturation accuracy in conditions of elevated methemoglobin(4 hours)
- Determine oxygen saturation accuracy in conditions of elevated carboxyhemoglobin(4 hours)
研究者
研究点 (1)
Loading locations...
相似试验
已完成
不适用
Dysfunctional Hemoglobin CO-oxElevated Levels of CarboxyhemoglobinNCT02427685Nonin Medical, Inc27
已完成
不适用
Dysfunctional Hemoglobin Pulse OximetryCarbon Monoxide PoisoningNCT03030833Nonin Medical, Inc36
已完成
不适用
Non-Invasive Measurement of Hemoglobin (Using Pulse Co-Oximetry) in Patients Undergoing Elective Cesarean DeliveryPostpartum HemorrhageNCT01108471Stanford University50
已完成
不适用
Validation of Noninvasive Hemoglobin, Comparison of PVI to Standard Methods, and Validation of Clinical Efficacy of ORISurgeryNCT02841397Masimo Corporation50
已完成
不适用
The Value of Continuous Noninvasive Hemoglobin Monitoring Using Masimo Radical -7 Pulse CO-Oximeter in Intraoperative Blood Transfusion Practice During Obstetric ProceduresMasimo Radical Pulse Co-oximeter in Intraoperative Blood Transfusion Practice During Obestatric ProcedureNCT03408938Cairo University60