Master Study for the MRI Compatibility of the EVIA/ENTOVIS Pacemaker in Combination With Safio S Pacemaker Lead

Biotronik SE & Co. KG9 个研究点 分布在 5 个国家目标入组 121 人开始时间: 2012年3月最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
121
试验地点
9
主要终点
Primary hypothesis 5: R-wave sensing attenuation

研究概览

简要总结

This investigation is designed to demonstrate the clinical safety of the EVIA/ ENTOVIS pacemaker system when used under specific MRI conditions.

详细描述

In the past, MR scans were always contraindicated for pacemaker patients. If particular prerequisites and conditions are fulfilled, MR scans can now be conducted on patients with BIOTRONIK's EVIA/ENTOVIS pacemakers in combination with the Safio S pacemaker leads, scheduled to be part of this clinical study.

The specific MRI conditions for this investigation with the EVIA/ENTOVIS pacemaker family and Safio S pacemaker leads are defined in the study protocol

研究设计

研究类型
Observational
观察模型
Case Only
时间视角
Prospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Pacemaker system consisting solely of the EVIA/ENTOVIS pacemaker and Safio S pacemaker leads
  • The pacemaker system has to be implanted, repositioned or exchanged at least 5 weeks prior to enrollment
  • Age more than 18 years
  • Understand the nature of the procedure
  • Able and willing to complete MRI testing
  • Able and willing to activate and use the Cardio Messenger
  • Give written informed consent
  • Able to complete all testing required by the clinical protocol
  • Ability to measure atrial and/or ventricular pacing threshold(s)* (at 0.4 ms)
  • All pacing thresholds do not exceed 2.0V @0.4ms.
  • Available for follow-up visit at the investigational site
  • Patient body height greater or equal to 140 cm
  • Pectoral implantation
  • The ascertained lead impedance is between 200 and 1500 Ohms.

排除标准

  • No EVIA/ENTOVIS /Safio S pacemaker system implanted
  • Pacing threshold(s) (at 0.4 ms) and sensing amplitudes are not measurable
  • Meet one or more of the contraindications
  • Being pregnant
  • Have a life expectancy of less than three months
  • Cardiac surgery already scheduled in the next three months
  • Enrolled in another cardiac clinical investigation
  • Have other medical implants that may interact with MRI, e.g. abandoned pacemaker/ICD leads, lead extensions, other active medical devices, non-MRI compatible devices (e.g. mechanical tricuspid valve)
  • Have other metallic artifacts/components in body that may interact with MRI

结局指标

主要结局

Primary hypothesis 5: R-wave sensing attenuation

时间窗: 1 month

R-wave attenuation: R-wave amplitude decrease (between pre-MRI follow-up and 1-month follow-up) exceeding 50% or a R-wave amplitude at 1-month follow-up smaller or equal to 5.0 mV.

Primary hypotheses 2 and 3: pacing threshold rise (atrial and ventricular)

时间窗: 1 month

The atrial or ventricular MRI induced pacing threshold rise is not larger than or equal to 0.5 V between Pre-MRI and 1-month follow-up

Primary hypothesis 4: P-wave sensing attenuation

时间窗: 1 month

P-wave attenuation: P-wave amplitude decrease (between pre-MRI follow-up and 1-month follow-up) exceeding 50% or a P-wave amplitude at 1-month follow-up smaller or equal to 1.5 mV

Primary hypothesis 1: MRI and pacing system related Serious Adverse Device Effect (SADE) free rate

时间窗: 3 months

MRI and pacing system related Serious Adverse Device Effect (SADE) Free-Rate is greater than 90%.

次要结局

未报告次要终点

研究者

申办方类型
Industry
责任方
Sponsor