NC Thermometer (Gen 3) ISO 80601-2-56 Validation

Natural Cycles2 个研究点 分布在 1 个国家目标入组 105 人开始时间: 2023年1月10日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
105
试验地点
2
主要终点
Clinical Bias

研究概览

简要总结

The purpose of this clinical validation is to evaluate the the clinical bias, the limits of agreement, and the clinical repeatability of measuring oral temperature using a new thermometer, NC° Thermometer (Gen 3), compared to a reference clinical thermometer.

详细描述

The purpose of this clinical study is to validate the clinical accuracy of the NC° Thermometer (Gen 3) according to the requirements of ISO 80601-2-56:2017. The NC° Thermometer (Gen 3) (the device under study (DUT)) is a predictive thermometer that operates in adjusted mode. The Welch Allyn oral thermometer (Welch Allyn Spot Vital Signs 420 SureTemp temperature module, Welch Allyn, San Diego, CA) with a measurement accuracy of ± 0.1 °C in monitor mode will be used the reference clinical thermometer (RCT). During the clinical study, subjects will be asked to perform three consecutive oral measurements with the DUT, and one oral measurement with the RCT.

研究设计

研究类型
Observational
观察模型
Other
时间视角
Other

入排标准

年龄范围
5 Years 至 —(Child, Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • They are more than 5 years old
  • They have not taken any antipyretic in the preceding 120 minutes
  • They have no medical conditions which might impact the clinical accuracy validation results, such as inflammation at the measuring site (the mouth)
  • They are not taking medications such as barbiturates, thyroid preparations, antipsychotics, and recent immunizations

排除标准

  • 未提供

结局指标

主要结局

Clinical Bias

时间窗: Up to 1 hour

Clinical bias is the mean difference between the output temperatures of the NC° Thermometer (Gen 3) (DUT) and the reference clinical thermometer (RCT) for a specific reference body site (the mouth) when measured from a selected group of subjects. The clinical bias of the DUT defines closeness between the DUT output temperature and that of the RCT.

Clinical repeatability

时间窗: Up to 1 hour

Clinical repeatability (sometimes called the perceived accuracy) is a measure of the consistency of repeated measurements with identical operating conditions when temperatures are taken within short time intervals from the same measuring site of the same subject by the same operator with the same intermittent clinical thermometer. Clinical repeatability is a pooled standard deviation of triplicate measurements over the population of subjects.

Limits of agreement

时间窗: Up to 1 hour

Dispersion of output temperatures around the clinical bias can be estimated by the standard deviation of the temperature differences between the DUT and the RCT as measured from multiple subjects.