临床试验/NCT00390702
NCT00390702已完成不适用

International Multicentre Clinical Device Investigation on Safety and Effectiveness of the Nit-Occlud® Lê VSD Spiral Coil System for VSD Occlusion Developed by Pfm AG, Cologne

pfm medical gmbh12 个研究点 分布在 2 个国家目标入组 103 人开始时间: 2006年10月最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
入组人数
103
试验地点
12
主要终点
a set of Performance Criteria (PC) deduced for VSD analogue to the set of Objective Performance Criteria (OPC) of the 'Multiorganization Advisory Panel to FDA for Pediatric Cardiovascular Devices' for PDAs

研究概览

简要总结

The Ventricular septal defect (VSD) is the most common of all congenital cardiac malformations. By modifying the Nit-Occlud® PDA Device the Nit-Occlud® Lê VSD Spiral System was designed. as a percutaneous, transcatheter device for occlusion of (peri)membranous and muscular ventricular septum defects (VSD) with a spiral coil. In this clinical investigation feasibility, safety and performance of the new cardiac occluder will be evaluated in accordance with European and US regulations. The study data will be compared to performance criteria for VSD, which are deduced analogue to the published specific Objective Performance Criteria (OPCs) for PDAs. The first part of the study has been performed in three clinical centres in Germany. For the second part in April 2009 four additional clinical centres in Germany, Israel, Italy and Spain were included.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Treatment
盲法
None

入排标准

年龄范围
2 Years 至 —(Child, Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • VSD must be diagnosed by acknowledged methods, like echocardiography
  • Signs of left ventricular volume overload are present (left ventricle or left atrium diameters 2 standard deviations greater than normal) and/or measured by catheterisation : Qp/Qs > 1.5
  • Pulmonary vascular resistance is less than 4 Wood units
  • The patient is older than 24 months
  • The VSD has a perimembranous or muscular location.
  • A distance between the rim of the VSD and the aortic annulus of at least 3.0 mm (as measured on the 2-D echocardiography - apical 5-chamber view)
  • The minimal diameter (size) of the VSD is less than 8.0 mm (as measured on the 2-D echocardiography - apical 5-chamber view)
  • Patient must agree to fully participate in the clinical trial and give informed consent in writing. If the patient is without legal ability additionally the person of legal responsibility must agree and s/he must give the informed consent in writing

排除标准

  • Pathological or physical condition precluding the implantation of a Nit-Occlud® Lê VSD coil, such as :
  • perimembranous VSD with no evidence of circular aneurysm formation
  • Associated cardiac anomalies requiring surgery (greater than mild aortic insufficiency; such as aortic valve prolapse)
  • Active endocarditis or other type of sepsis or other active infection at time of implantation
  • Thrombus at or near the intended site of implantation
  • Thrombus in the vessels through which access to the VSD is gained (unless the patient is protected with an embolic protection device such as a vena cava filter)
  • Vessels through which access to the VSD is gained can not accommodate a 7 F sheath
  • Potential steric (3-dimensional) interference of the occluder with intracardiac or intravascular structures (like valves)
  • History of blood disorder (coagulopathy, tendency towards haemolysis)
  • History of hypersensitivity to contrast medium or Nitinol
  • AV-block II° or III°, atrial fibrillation, or atrial flutter
  • End stage cardiac disease, irreversible major organ failure, or terminal cancer
  • HIV infection
  • Cerebrovascular disease or neurological deterioration
  • Emergency cardiologic intervention
  • Patient*) is not able to fully participate in this study including all follow-ups (e.g. for mental or geographical reasons, or patient is intravenous drug user or has strong potential for non-compliance to medical regimes)
  • Patient, respectively the person of legal responsibility, is mentally unable to understand the nature, aims, or possible consequences of the clinical investigation
  • Pregnant or breast-feeding women
  • Patient did participate in another clinical investigation during the last 3 months
  • Patient or the person of responsibility has revoked the consent.

研究组 & 干预措施

VSD occluder

Experimental

transcatheter implantation of a VSD occluder (Nitinol coil)

干预措施: transcatheter implantation of a VSD occluder (Nitinol coil) (Device)

结局指标

主要结局

a set of Performance Criteria (PC) deduced for VSD analogue to the set of Objective Performance Criteria (OPC) of the 'Multiorganization Advisory Panel to FDA for Pediatric Cardiovascular Devices' for PDAs

时间窗: Five years

次要结局

未报告次要终点