跳至主要内容
临床试验/NCT01843270
NCT01843270已完成不适用

Appropriate Laryngeal Mask Airway Size in Patients With Overweight: A Comparison of the Actual Body Weight and Ideal Body Weight

Yonsei University2 个研究点 分布在 1 个国家目标入组 100 人开始时间: 2013年11月最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
100
试验地点
2
主要终点
Oropharyngeal leak pressure

研究概览

简要总结

The classic laryngeal mask airway is one of the keystones of modern anaesthetic practice. It achieves a reliable airway with a low incidence of major complications and pharynges-laryngeal morbidity. The manufactures of laryngeal mask airway recommend the proper size selection by actual body weight. It was recommended a size 3 for 30-70kg, a size 4 for 70-90 kg, and size 5 for > 90 kg. Obesity increases the fat tissue around an upper airway and decreases the pharyngeal cross sectional area. Also it was reported that pharyngeal area was reduced according to the increasing of BMI. Therefore, the aim of this study is to determine which of actual body weight or ideal body weight is proper on the size selection of laryngeal mask airway.

研究设计

研究类型
Interventional
分配方式
Randomized
干预模型
Parallel
主要目的
Prevention
盲法
None

入排标准

年龄范围
20 Years 至 70 Years(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • age 20~70
  • body mass index > 25 kg/m2
  • patient who needs surgery with laryngeal mask airway

排除标准

  • difficult airway
  • upper airway infection or disease of airway
  • history of gastroesophageal reflux
  • history of head and neck surgery

结局指标

主要结局

Oropharyngeal leak pressure

时间窗: 10mins

Oropharyngeal leak pressure is measured when the airway pressure becomes the plateau state in condition of gas flow 3 L/min and flow out valve 30 cmH2O.

次要结局

未报告次要终点

研究者

申办方类型
Other
责任方
Sponsor

研究点 (2)

Loading locations...

相似试验