Pacing Capture Threshold Change Following 3T MRI Scan of MR Conditional Cardiac Implantable Electronic Devices

Medtronic0 个研究点目标入组 41 人开始时间: 2016年4月最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
Medtronic
入组人数
41
主要终点
The proportion of patients implanted with an MR Conditional CIED System with a pre-MRI to post-MRI scan PCT measure change of >0.5 Volts following 3T MRI scan exposure.

研究概览

简要总结

The purpose of this study is to evaluate product performance of Magnetic Resonance (MR) Conditional Cardiac Implantable Electronic Devices (CIED) following 3 tesla (3T) MRI exposure. This will be achieved by evaluating the changes in pacing capture threshold (PCT) measurements following 3T MRI scan exposure. This study is required by FDA as a condition of approval of 3T MRI compatible labeling of applicable CIED systems. This study is conducted within Medtronic's post-market surveillance platform, the Product Surveillance Registry (PSR).

研究设计

研究类型
Observational
观察模型
Other
时间视角
Prospective

入排标准

性别
All
接受健康志愿者

入选标准

  • Subject or legally authorized representative provides written authorization and/or consent per institution and geographical requirements
  • Subject is intended to be implanted or is within 30 days post-implant of a qualifying Medtronic 3T MR Conditional CIED system

排除标准

  • Subject who is, or will be inaccessible for follow-up at a study site
  • Subject with exclusion criteria required by local law (Europe, Central Asia [ECA] only)
  • Implant and follow-up data, including any adverse device effects and system modifications at implant through the time of enrollment are unavailable

结局指标

主要结局

The proportion of patients implanted with an MR Conditional CIED System with a pre-MRI to post-MRI scan PCT measure change of >0.5 Volts following 3T MRI scan exposure.

时间窗: Patients will be followed for approximately 5 years.

次要结局

未报告次要终点

研究者

发起方
Medtronic
申办方类型
Industry
责任方
Sponsor