Method and Sample Comparison Study of Minicare BNP at a Central Laboratory

Philips Handheld Diagnostics1 个研究点 分布在 1 个国家目标入组 296 人开始时间: 2017年2月27日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
296
试验地点
1
主要终点
To investigate the bias between the Philips Minicare BNP test with the Centaur BNP test of the measurement procedures.BNP level measured in human blood samples (EDTA venous whole blood and/or EDTA plasma) will be assessed.

研究概览

简要总结

Sample Comparison The objective of this study is to compare the Philips Minicare BNP test with the Centaur BNP test. BNP level measured in human blood samples (EDTA venous whole blood and/or EDTA plasma) will be assessed. Matrix Comparison To compare BNP measurements obtained using Minicare System for EDTA whole blood and EDTA plasma samples.Prior to the study start, a proficiency study will be performed on the Minicare BNP, Architect and I-STAT BNP assay to be sure that devices are performing according to expectations. Approximately 100 EDTA patient whole blood samples covering the entire measuring range of Minicare BNP from LOQ until end of measuring range (expected to be approximately 5000 ng/L) (Ref 7) will be tested (according to CLSI EP09A3 (Ref 5)) at a central laboratory using the Minicare BNP, Alere BNP, I-STAT BNP system. The outcome of the site routine assay (Centaur BNP systems) will be used to evaluate if the blood samples cover the entire distribution range of the assay. If necessary, up to 20% of the samples may be spiked so that the entire range of the assay is covered. The central laboratory will identify the blood tubes with a BNP request. Approximately 1000 μl EDTA whole blood will be available (left over material) for testing on the non-routine devices (i.e., Minicare BNP, Alere BNP, and I-STAT BNP). Testing for the Alere BNP, I-STAT BNP, and Centaur BNP systems will be performed according to the respective IFU's (Ref 1, 2, 3, 4). For the Minicare BNP system, test will be performed according to a DRAFT IFU or a similar document describing the sample treatment and handling for the Minicare BNP system. Alere BNP, I-STAT BNP and Minicare BNP samples should be tested within 6 hours after collection and Centaur BNP samples within 8h of collection. Although different stability claims for BNP, it is highly recommended to perform all measurements on all systems within 2 hours. For Alere BNP, I-STAT BNP, Centaur BNP analysis will be performed using a single blood sample for each patient. Alere BNP and I-STAT will use EDTA whole blood samples as sample type Centaur BNP will be tested on EDTA plasma samples. For the Minicare BNP, both EDTA whole blood and EDTA plasma will be tested for each patient. For the Minicare BNP system the whole blood samples will be tested using a pipette and the same sample will be tested using the SmearSafe Blood Dispenser (transfer device) to investigate a reliable use of such transfer device.

Only the routine Centaur BNP outcomes will be used as part of support for patient diagnosis.

研究设计

研究类型
Observational
观察模型
Other
时间视角
Prospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • 未提供

排除标准

  • 未提供

结局指标

主要结局

To investigate the bias between the Philips Minicare BNP test with the Centaur BNP test of the measurement procedures.BNP level measured in human blood samples (EDTA venous whole blood and/or EDTA plasma) will be assessed.

时间窗: May 2017

To investigate the bias between the Philips Minicare BNP test with the Centaur BNP test using values spanning the common measuring interval of the measurement procedures.BNP level measured in human blood samples (EDTA venous whole blood and/or EDTA plasma) will be assessed.

To compare BNP measurements obtained using Minicare System for EDTA whole blood and EDTA plasma samples.

时间窗: May 2017

To compare BNP measurements obtained using Minicare System for EDTA whole blood and EDTA plasma samples.

次要结局

  • Percentage of agreement systems(May 2017)
  • Percentage of agreement samples(May 2017)
  • Evaluate performance of the SmearSafe Blood Dispenser (transfer device), for transferring whole blood from the blood tube directly to the Minicare BNP cartridge.(May 2017)

研究者

发起方
Philips Handheld Diagnostics
申办方类型
Industry
责任方
Sponsor