The Effect of Electrical Stimulation on Impairment of the Painful Post-Stroke Shoulder

MetroHealth Medical Center0 个研究点目标入组 10 人开始时间: 2010年12月27日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
入组人数
10
主要终点
Passive Range of Motion of Shoulder in Shoulder External Rotation

研究概览

简要总结

Examination of the effects of short-term use of TENS and t-NMES on passive pain-free ROM of the painful post-stroke shoulder. Treatments are compared with no stimulation.

详细描述

Single cohort, cross-over study. Subjects undergo passive pain-free ROM of the painful post-stroke shoulder under 3 stimulation conditions. The stimulation was delivered by a trained occupational therapist under three conditions: 10 seconds of TENS, 10 seconds of t-NMES, and 10 seconds of no stimulation. Each subject was exposed to each of the three stimulation conditions three times in a computer-generated random sequence for each outcome measure with a 5-minute wash-out period between each stimulation. This protocol is repeated for each of the two passive movements tested, shoulder abduction and shoulder external rotation.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Treatment
盲法
None

盲法说明

Participant, Investigator, and Outcomes Assessor were masked to the particular stimulation during each range of motion assessment

入排标准

年龄范围
21 Years 至 89 Years(Adult, Older Adult)
性别
All
接受健康志愿者
否

入选标准

  • •at least 3 months post-stroke
  • •worst shoulder pain in the last week at least 4 on the 0-10 numerical rating scale,
  • •adequate cognitive ability to be able to rate their pain in the past week

排除标准

  • •history of tachyarrhythmia with decreased blood pressure
  • •uncontrolled seizures (defined as more than one per month)
  • •an implanted electrical device
  • •uncompensated hemi-neglect

研究组 & 干预措施

TENS/t-NMES/No stimulation

Experimental

TENS stimulation parameters were of a symmetric waveform, a frequency of 100 Hz, and a pulse duration of 300 microseconds (EMPI 300PVTM program PPR #7) applied for 10 seconds. t-NMES parameters were a symmetric waveform with 2 second ramp-up and 2 second ramp-down, a frequency of 35Hz, and a pulse duration of 300 microseconds (EMPI 300PVTM program PPR #3). The t-NMES current intensity was set by adjusting the amplitude to yield the strongest contraction of the underlying muscles without initiating pain. Device and electrodes remained in place but no stimulation was delivered over the 10 second interval. Exposed to each stimulation 3 times for each shoulder ROM.

干预措施: t-NMES (Device)

TENS/t-NMES/No stimulation

Experimental

TENS stimulation parameters were of a symmetric waveform, a frequency of 100 Hz, and a pulse duration of 300 microseconds (EMPI 300PVTM program PPR #7) applied for 10 seconds. t-NMES parameters were a symmetric waveform with 2 second ramp-up and 2 second ramp-down, a frequency of 35Hz, and a pulse duration of 300 microseconds (EMPI 300PVTM program PPR #3). The t-NMES current intensity was set by adjusting the amplitude to yield the strongest contraction of the underlying muscles without initiating pain. Device and electrodes remained in place but no stimulation was delivered over the 10 second interval. Exposed to each stimulation 3 times for each shoulder ROM.

干预措施: No stimulation (Other)

TENS/t-NMES/No stimulation

Experimental

TENS stimulation parameters were of a symmetric waveform, a frequency of 100 Hz, and a pulse duration of 300 microseconds (EMPI 300PVTM program PPR #7) applied for 10 seconds. t-NMES parameters were a symmetric waveform with 2 second ramp-up and 2 second ramp-down, a frequency of 35Hz, and a pulse duration of 300 microseconds (EMPI 300PVTM program PPR #3). The t-NMES current intensity was set by adjusting the amplitude to yield the strongest contraction of the underlying muscles without initiating pain. Device and electrodes remained in place but no stimulation was delivered over the 10 second interval. Exposed to each stimulation 3 times for each shoulder ROM.

干预措施: TENS (Device)

结局指标

主要结局

Passive Range of Motion of Shoulder in Shoulder External Rotation

时间窗: 10 seconds

passive range of motion of shoulder in shoulder external rotation as measured with hand-held goniometer

Passive Range of Motion of Shoulder in Shoulder Abduction

时间窗: 10 seconds

passive range of motion of shoulder in shoulder abduction as measured with hand-held goniometer

次要结局

未报告次要终点

研究者

申办方类型
Other
责任方
Principal Investigator
主要研究者

John Chae, MD

VP Research Instititute; Chair, Department of PM&R

MetroHealth Medical Center