临床试验/NCT01441583
NCT01441583已完成不适用

Ingenio Device Algorithm Study

Boston Scientific Corporation41 个研究点 分布在 12 个国家目标入组 139 人开始时间: 2011年10月最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
入组人数
139
试验地点
41
主要终点
Accuracy of Commanded Right Atrial Automatic Threshold (RAAT)

研究概览

简要总结

The purpose of the IVORY Study is to gather data to support global submissions/approvals for some models of the Ingenio device family.

详细描述

IVORY is a prospective, multi-center, randomized within-patient, single-blinded study to gather data to support Right Atrial Auto Threshold and RYTHMIQ

研究设计

研究类型
Interventional
分配方式
Randomized
干预模型
Crossover
主要目的
Treatment
盲法
Single (Participant)

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Subjects who are willing and capable of providing informed consent to undergo a device implant and to participate in all testing associated with this clinical study;
  • Subjects whose age is 18 or above, or of legal age to give informed consent specific to national law;
  • Subjects indicated for a dual chamber pacemaker or a CRT-P device according to class I or class II indications of the standard ESC or ACC / AHA implant guidelines;
  • Subjects who are planned to be implanted with all leads intended for a specific device type (dual chamber pacemaker: atrial and right ventricular lead, CRT-P: atrial, right and left ventricular lead) or are already implanted with such leads;
  • Subjects who receive or are implanted with a bipolar atrial lead.

排除标准

  • Women of childbearing potential who are or might be pregnant at the time of the study (method of assessment upon physician's discretion);
  • Enrolled in any other concurrent study, with the exception of local mandatory governmental registries and observational studies/registries that are not in conflict or affect
  • Schedule of procedures for IVORY (i.e. should not cause additional or missed visits);
  • Programming of devices for IVORY per CIP;
  • IVORY outcome (i.e. involve medications that could affect the heart rate of the subject);
  • Conduct of IVORY per GCP / ISO 14 155:2011 / local regulations.
  • Subjects who live at such a distance from the clinic that travels for FU visits at a study center would be unusually difficult or burdensome for the subject;
  • Inability or refusal to comply with the FU schedule;
  • A life expectancy of less than 12 months, per physician discretion;
  • Subjects who are planned to be programmed to a pacing mode other than DDD / DDDR during the study period.

研究组 & 干预措施

Pacemaker - RYTHMIQ Off at Pre-discharge, On at 1-Month

Active Comparator

For the RYTHMIQ endpoint only, subjects implanted with a pacemaker will be randomized. Those randomized to RYTHMIQ Off at Pre-Discharge will have RYTHMIQ programmed Off until their 1-month follow up, when they will be crossed over to RYTHMIQ On until their 3-month follow up.

干预措施: RAAT (Device)

Pacemaker - RYTHMIQ Off at Pre-discharge, On at 1-Month

Active Comparator

For the RYTHMIQ endpoint only, subjects implanted with a pacemaker will be randomized. Those randomized to RYTHMIQ Off at Pre-Discharge will have RYTHMIQ programmed Off until their 1-month follow up, when they will be crossed over to RYTHMIQ On until their 3-month follow up.

干预措施: RYTHMIQ (Device)

Pacemaker - RYTHMIQ On at Pre-Discharge, Off at 1-Month

Active Comparator

For the RYTHMIQ endpoint only, subjects implanted with a pacemaker will be randomized. Those randomized to RYTHMIQ On at Pre-Discharge will have RYTHMIQ programmed On until their 1-month follow up, when they will be crossed over to RYTHMIQ Off until their 3-month follow up.

干预措施: RAAT (Device)

Pacemaker - RYTHMIQ On at Pre-Discharge, Off at 1-Month

Active Comparator

For the RYTHMIQ endpoint only, subjects implanted with a pacemaker will be randomized. Those randomized to RYTHMIQ On at Pre-Discharge will have RYTHMIQ programmed On until their 1-month follow up, when they will be crossed over to RYTHMIQ Off until their 3-month follow up.

干预措施: RYTHMIQ (Device)

CRT-P

Experimental

CRT-P devices were not involved in RYTHMIQ endpoint evaluation, only RAAT.

干预措施: RAAT (Device)

结局指标

主要结局

Accuracy of Commanded Right Atrial Automatic Threshold (RAAT)

时间窗: 3 months post implant

Accuracy was evaluated by the proportion of tests resulting in an RAAT commanded threshold with \|Manual unipolar threshold - commanded threshold\| ≤ 0.5V. Subjects were allowed to contribute multiple paired datasets for this endpoint analysis, one set each from the 1-month and 3-month visits. Paired datasets from the 1-month and 3-month visits were pooled for purposes of endpoint analysis. A paired dataset consisted of the manual unipolar threshold and the commanded threshold obtained at a visit.

System-related Complication-free Rate

时间窗: 90 days post-implant

Arms/Groups are only applicable for the outcome measure 'Secondary Outcome: RYTHMIQ - Median Relative Reduction of RV Pacing Percent (RVPP) for RYTHMIQ Programmed to ON vs. RYTHMIQ Programmed to OFF'. This outcome measure was analyzed only for the combined population and both pacemakers and CRT-P devices were pooled for the analysis. The analysis assesses the system-related complication (SRC) -free rate (excluding LV related events for CRT-P subjects) for subjects implanted or attempted to be implanted with a study device through the first 90 days following the implant procedure.

次要结局

  • RYTHMIQ - Median Relative Reduction of RV Pacing Percent (RVPP) for RYTHMIQ On at Pre-Discharge, Off at 1-Month vs. RYTHMIQ Off at Pre-discharge, On at 1-Month(3 months post implant)
  • Accuracy of Ambulatory RAAT(3 months post-implant)
  • Appropriate RAAT Test Outcome(3 months post-implant)
  • Percentage of RAAT Threshold Tests With Sufficient RAAT Pace Output Margin(3 months post implant)