Evaluation of the CNOGA TensorTip COMBO GLUCOMETER (CoG) During Standardized Meal Test and Consecutive Home Use by Patients With Diabetes Mellitus
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 40
- 试验地点
- 1
- 主要终点
- CoG accuracy in monitoring glucose levels (invasively) during a standardized meal experiment.
研究概览
简要总结
The CNOGA Combo Glucometer (CoG) employs an invasive glucose meter using blood glucose teststrips and requiering a capillary blood sample obtained from a fingerstick and in addition, a non-invasive optical component for prediction of tissue glucose at the fingertip. In this study, participants will use both devices during meal tests and also at home during routine care. The results from the device will be compared to a standard reference method and will be used to determine the accuracy of the two device components.
详细描述
The trial starts with a screening and training visit (V1). This is followed by an individual calibration period for the non-invasive device component by means of multiple parallel measurements with the invasive and the noninvasive device component. This will be done by the participants at home. Thereafter, the participants will come to the study site and will ingest a standardized meal (V2). Blood will be drawn for assessment of HbA1c and biochemical safety parameters. During the next 3 hours, 11 fingersticks will be made to obtain capillary blood samples for comparator readings with the two CoG components and with a laboratory reference method (YSI gucose analyzer). The patients will now continue to use the devices at home for three months, but are not allowed to use any CoG result for treatment decisions. In an interim visit after 6 weeks (V3), blood will be drawrn for HbA1c and biochemical safety parameters and three comparison measurements will be made with the CoG device components and the laboratory reference method. At the final visit after 12 weeks (V4), the procedures of the first meal test visit (V2) will be repeated, which conludes that study. In addition, the participants will be asked to complete a device satisfaction questionnaire at V2, V3, and V4.
研究设计
- 研究类型
- Observational
- 观察模型
- Cohort
- 时间视角
- Prospective
入排标准
- 年龄范围
- 18 Years 至 90 Years(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Patients with Type 1 or Type 2 diabetes of Afro-American, Hispanic, or Asian-American origin
- •Subjects who are able to provide informed consent (by him/herself or by his/her guardian);
- •18 years old and above;
- •HbA1c < 10 %
- •Anatomically suitable finger as determined by the clinical investigator and detailed in section 4
排除标准
- •Does not meet inclusion criteria;
- •Subjects requiring dialysis;
- •Any conditions that may hamper good visual contact between the finger and sensor, such as raised birthmarks, scars and/or tattoos;
- •Pregnancy;
- •Nursing mothers;
- •Any skin injuries on the measured finger;
- •Severe disease conditions (cancer etc.)
- •Any condition that may prevent patients from successful participation in the trial (in discretion of the investigator)
结局指标
主要结局
CoG accuracy in monitoring glucose levels (invasively) during a standardized meal experiment.
时间窗: Experiment will be performed twice within 3 months
Comparison of CoG invasive device readings (blood glucose in mg/dL) with the results of a standard laboratory reference method (YSI glucose analyzer, blood glucose in mg/dL) by consensus-error grid analysis
CoG accuracy in measuring glucose levels (non-invasively) during a standardized meal experiment.
时间窗: Experiment will be performed twice within 3 months
Comparison of non-invasive CoG device measurements (tissue glucose in mg/dL) with the results of a standard laboratory reference method (YSI glucose analyzer, blood glucose in mg/dL) by consensus-error grid analysis
次要结局
- Impact of the CoG on long-term (three month) glycemic control (HbA1c using standard laboratory procedures) in patients with type 1 and 2 diabetes(3 months)
- Treatment satisfaction questionnaire(3 months)