临床试验/NCT05635513
NCT05635513已完成不适用

Owlet Smart Sock 3 Accuracy on Infants

Owlet Baby Care, Inc.1 个研究点 分布在 1 个国家目标入组 28 人开始时间: 2021年7月14日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
28
试验地点
1
主要终点
Owlet Oxygen Saturation Accuracy Performance

研究概览

简要总结

The purpose of this research study is to evaluate the Owlet Smart Sock 3 wireless pulse oximeter performance in assessing the heart rate and oxygen saturation in neonatal population, while measuring the same parameter with a standard wired pulse oximeter simultaneously. The monitoring duration with the study device will be a maximum of 60 minutes, after which the recorded data from the Owlet Smart Sock 3 and the standard wired pulse oximeters will be de-identified and analyzed.

详细描述

This prospective observational study compared the HR and SpO2 measurements of the OSS3 (Owlet, Utah, USA) with the Masimo SET® (Masimo Corporation, California, USA) in infants admitted to the University of Utah NICU. The inclusion criteria were infants who weighed 1.7-2.5 kg at the time of monitoring and were medically stable to tolerate routine care. The exclusion criteria were: infants with significant anomalies, congenital deformities, or medical instability. Infants with an intravenous catheter (IV) and skin lesions on either foot were excluded from the study. The wireless device being tested was the OSS3. The gold standard for validating the pulse oximeter's accuracy is arterial blood oxygen saturation (SaO 2 ) at the pulse oximeter sensor site. [6] However, obtaining SaO2 is an invasive arterial blood test for infants. To avoid this conflict, the OSS3 was tested against the reference FDA-approved wired pulse oximeter Masimo.

The study was approved by the University of Utah Institutional Review Board, and informed consent was obtained from parents. During the measurement, OSS3 and Masimo were placed on the infants with one device on each foot. Since the data obtained from the study devices can not be recorded in the hospital electronic medical record, the hospital wired pulse oximeter (Philips, Amsterdam, Netherlands) continued to be used on one of the infant's hands during the study period. OSS3 and Masimo were calibrated appropriately against the hospital wired pulse oximeter with matching HR and SpO 2 readings. Pulse oximetry data were collected simultaneously for 60 minutes on the quiet infants. The two devices (OSS3 and Masimo) continuously recorded HR (beats per minute, bpm) and SpO2 (%) every two seconds. Data from the 2 devices were uploaded and compared according to their time stamped measurements. The paired data agreement was compared using Pearson's correlation coefficient (r), Bland-Altman, Average Root Mean Square (ARMS), and Prevalence and Bias Adjusted Kappa (PABAK) analyses.

研究设计

研究类型
Observational
观察模型
Other
时间视角
Prospective

入排标准

性别
All
接受健康志愿者

入选标准

  • Infant who weigh between 1.7kg and 2.3kg
  • Infant who is admitted to U of U NICU

排除标准

  • Infant who has injuries, deformities, abnormalities, or intravenous catheter in the extremities that may prevent proper application of device
  • Infant who has unstable vital signs or clinical condition
  • No parental consent
  • Inability to achieve adequate fit with the OSS3

结局指标

主要结局

Owlet Oxygen Saturation Accuracy Performance

时间窗: During the monitoring period (approx. 1 hour)

Conduct a single-arm observational study to monitor oxygen saturation (SpO2) while a standard wired pulse oximeter (WPOx) will simultaneously evaluate the same parameters in the University of Utah (U of U) Neonatal Intensive Care Unit (NICU).

Owlet Pulse Rate Accuracy Performance

时间窗: During the monitoring period (approx. 1 hour)

Conduct a single-arm observational study to monitor heart rate (HR), while a standard wired pulse oximeter (WPOx) will simultaneously evaluate the same parameters in the University of Utah (U of U) Neonatal Intensive Care Unit (NICU).

次要结局

未报告次要终点