跳至主要内容
临床试验/JPRN-UMIN000005287
JPRN-UMIN000005287已完成未知

Comparative study of human choroidal structure by different OCT machines - Choroidal structure by OCT

Ophthalmology, Kagoshima Univ.0 个研究点目标入组 40 人开始时间: 2011年3月22日最近更新:
适应症

试验速览

阶段
未知
状态
已完成
发起方
入组人数
40

研究概览

简要总结

Results:Forty-three subjects (mean age 30.5 years) were enrolled. Of 43 the eyes, SCT of 39 eyes (90.7%) could be measured using each machine. Intra-class correlation coefficients (95% confident interval) were 0.976 (0.954-9.987), 0.958 (0.919-0.978) and 0.939 (0.895-0.971) with Spectralis, Cirrus, and Topcon respectively. Inter-rater correlation coefficients (95% confident interval) were 0.944 (0.893 to 0.971), 0.956 (0.831 to 0.983) and 0.924 (0.825 to 0.964) with Spectralis, Cirrus, and Topcon respectively. The average SCT was 272.6, 272.8, and 269.2 &micro;m with Spectralis, Cirrus, and Topcon, respectively. Inter-machine correlation coefficient was significantly high among the machines (P<0.001, Spearman); 0.97 (Spectralis-Cirrus), 0.96 (Cirrus-Topcon) and 0.98 (Topcon-Cirrus). Bland-Altman plot analysis showed no typical trend among the machines.Conclusions:SCT measurements obtained with 3 different SD-OCTs were highly correlated and could be used interchangeably.

研究设计

研究类型
Observational

入排标准

年龄范围
18years-old 至 40years-old(—)
性别
All

入选标准

  • 未提供

排除标准

  • Highly myopic eyes -5D or greater Those of which fundus cannot be observed. Those with apparent ocular diseases such as glaucoma, diabetic retinopathy, macular degenerations and others.

研究者

发起方
Ophthalmology, Kagoshima Univ.

相似试验

Comparative study of human choroidal structure by... | 临床试验