Validation of a Diagnostic Test for Glucose-6-phosphate Dehydrogenase Deficiency: Diagnostic Accuracy and Repeatability in Capillary Samples
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 229
- 试验地点
- 2
- 主要终点
- Sensitivity of SD Biosensor STANDARD G6PD Test for Identifying Women With Intermediate G6PD Activity
研究概览
简要总结
The primary objective of the study is to assess the accuracy of the SD Biosensor STANDARD™ point-of-care (POC) G6PD test in measuring G6PD activity and classifying results compared to a reference assay and across repeated measurements in capillary samples.
详细描述
Cross-sectional diagnostic accuracy study with 250 volunteer participants. The clinic will recruit and consent adult study participants. Clinic staff will draw venous blood and obtain finger stick capillary blood samples. Clinic staff will perform the investigational SD Biosensor STANDARD™ G6PD test for glucose-6-phosphate dehydrogenase (G6PD) deficiency and a HemoCue® hemoglobin test on finger stick capillary blood. An anti-coagulated venous blood sample will be sent to a Clinical Laboratory Improvement Amendments (CLIA) certified lab for G6PD reference testing by the gold standard assays: G6PD measurement by spectrophotometry using the Pointe Scientific G6PD reference assay and hemoglobin measurement by a hematology analyzer.
Individuals identified as G6PD deficient or intermediate by the reference test will be notified of their results by the clinic and referred to their physician for follow-up.
This study includes a nested repeatability study and a nested sample stability study. In the nested repeatability study up to 20 consented participants will provide 4 additional finger stick samples. Clinic staff will perform the SD Biosensor STANDARD™ G6PD test to assess the repeatability of the test in capillary blood over 8 G6PD and hemoglobin measurements on up to 3 different instruments.
In the nested sample stability study up to 8 consented participants will provide an additional venous blood draw sample to be tested at additional time points to document stability of the samples over time when tested by the SD Biosensor STANDARD™ G6PD Test.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Diagnostic
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 65 Years(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •Must communicate an understanding of the study protocol.
- •Must be able to provide written consent to undergo screening and provide medical history.
- •Participants must be afebrile and in general good health in the opinion of the investigator as determined by vital signs, medical history, and physical examination.
- •Black/African-American, by self-report.
排除标准
- •Blood transfusion in the past 3 months by self-report
结局指标
主要结局
Sensitivity of SD Biosensor STANDARD G6PD Test for Identifying Women With Intermediate G6PD Activity
时间窗: All samples were collected on study day 1
To investigate the performance of the SD Biosensor STANDARD G6PD test to distinguish females with intermediate G6PD activity from females with normal activity, assay sensitivity was determined at a G6PD activity threshold of 70%. A true positive for intermediate G6PD activity in females was defined as G6PD activity between 30 and 70% of normal in circulating venous blood as determined by the Pointe Scientific test. Sensitivity of the SD Biosensor STANDARD G6PD test is the percentage of women who tested positive for intermediate G6PD activity on the reference test who were identified by the test assay as positive, calculated as: Number of women who were both test and true positive / (Number of women who were test and true positive + Number of women who were test negative but true positive \[ie false negative\]) \* 100%. Sensitivity of the SD Biosensor STANDARD G6PD test for identifying females with intermediate G6PD activity was calculated from both venous and capillary blood samples.
Sensitivity of SD Biosensor STANDARD G6PD Test for Identifying G6PD Deficient Individuals
时间窗: All samples were collected on study day 1
For the purposes of this study, an individual was considered G6PD deficient if they tested positive in the Pointe Scientific assay. A true positive was defined as a participant with ≤ 30% of normal G6PD activity in circulating venous blood determined by the Pointe Scientific test. Diagnostic sensitivity of the SD Biosensor STANDARD G6PD test is defined as the percentage of participants who tested positive for G6PD deficiency on the reference test who were identified by the test assay as positive, calculated as: Number of participants who were both test and true positive / (Number of participants who were test and true positive + Number of participants who were test negative but true positive \[ie false negative\]) \* 100%. Sensitivity of the SD Biosensor STANDARD G6PD test was calculated from both venous and capillary blood samples.
Specificity of SD Biosensor STANDARD G6PD Test for Identifying G6PD Deficient Individuals
时间窗: All samples were collected on study day 1
For the purposes of this study, an individual was considered G6PD deficient if they tested positive by the Pointe Scientific assay. A true negative was defined as \> 30% of normal G6PD activity in circulating venous blood as determined by the Pointe Scientific test. Specificity is the percentage of participants who tested negative for G6PD deficiency on the reference test who were identified as negative using the test assay, calculated as: Number of participants who were both test and true negative / (Number of participants who were test and true negative + Number of participants who were test positive but true negative \[ie false positive\]) \* 100%. Specificity of the SD Biosensor STANDARD G6PD test was calculated from both venous and capillary blood samples.
Specificity of SD Biosensor STANDARD G6PD Test for Identifying Women With Intermediate G6PD Activity
时间窗: All samples were collected on study day 1
To investigate the performance of the SD Biosensor STANDARD G6PD test to distinguish females with intermediate G6PD activity from females with normal activity, the specificity was determined at a G6PD activity threshold of 70%. A true negative for intermediate G6PD activity in females was defined as G6PD activity \> 70% of normal in circulating venous blood as determined by the Pointe Scientific test. Specificity is the percentage of participants who tested negative for G6PD deficiency on the reference test who were identified as negative using the test assay, calculated as: Number of participants who were both test and true negative / (Number of participants who were test and true negative + Number of participants who were test positive but true negative \[ie false positive\]) \* 100%. Specificity of the SD Biosensor STANDARD G6PD test was calculated from both venous and capillary blood samples.
次要结局
- Repeatability Substudy: G6PD Levels in Capillary Blood Samples Measured Using the SD Biosensor G6PD Test, by Operator and Analyzer(All samples were collected on study day 1 and tested immediately)
- Repeatability Substudy: Hemoglobin Levels in Capillary Blood Samples Measured Using the SD Biosensor STANDARD Hemoglobin Test, by Operator and Analyzer(All samples were collected on study day 1 and tested immediately)
- Sample Stability Substudy: G6PD Levels Measured Over Time After Storage at Room Temperature(Blood samples were collected on day 1. Samples were tested immediately (< 1 hour) and at 2, 4, 6, and 24 hours after collection.)
- Sample Stability Substudy: Hemoglobin Levels Measured Over Time After Refrigerated Storage(Blood samples were collected on day 1. Samples were tested immediately (< 1 hour) and at 24, 48, 72, 96, and 168 hours after collection.)
- Accuracy Between the SD Biosensor STANDARD G6PD Test Measure of Hemoglobin and the Reference Hemoglobin Test(All samples were collected on study day 1)
- Sample Stability Substudy: G6PD Levels Measured Over Time After Refrigerated Storage(Blood samples were collected on day 1. Samples were tested immediately (< 1 hour) and at 24, 48, 72, 96, and 168 hours after collection.)
- Sample Stability Substudy: Hemoglobin Levels Measured Over Time After Storage at Room Temperature(Blood samples were collected on day 1. Samples were tested immediately (< 1 hour) and at 2, 4, 6, and 24 hours after collection.)
- Accuracy Between the SD Biosensor STANDARD G6PD Test Assay and the Pointe Scientific Test Kit(All samples were collected on study day 1)