Simulation of Stent-graft Deployment in Aortic Arch Aneurysms
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 4
- 试验地点
- 2
- 主要终点
- In vitro versus stimulation test
研究概览
简要总结
Total endovascular repair of the aortic arch represents a promising option for patients ineligible to open surgery. Custom-made design of stent-grafts (SG), such as the Terumo Aortic® RelayBranch device (DB), requires complex preoperative measures. Accurate SG deployment is required to avoid intraoperative or postoperative complications, which is extremely challenging in the aortic arch. In that context, the investigators aim is to develop a computational tool able to predict SG deployment in such highly complex situations. Four patient-specific cases will be performed with complete deployment of the DB and its bridging stents in aneurysmal aortic arch. Deviations of simulation predictions from actual stent positions will be estimated based on post-operative scan and a sensitivity analysis will be performed to assess the effects of material parameters. If good agreement between simulation and reality is obtained, numerical simulations will show their ability to successfully predict the DB deployment in complex anatomy. The results will emphasize the potential of computational simulations to assist practitioners in planning and performing complex and secure interventions.
详细描述
The process of numerical simulation of double-branched device in aortic arch aneurysm involves: (i) aortic arch segmentation with data from the preoperative CT-scan (ii) numerical modeling of the actual stentgraft used to treat the patient (iii) simulation of stentgraft deployment in aortic arch (iv) validation of simulation results based on post-operative CT scan images The goal of this study is to develop a numerical model of stentgraft deployment able to predict the actual deployement of the double branched device in aortic arch aneurysm.
研究设计
- 研究类型
- Observational
- 观察模型
- Cohort
- 时间视角
- Retrospective
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Aortic arch aneurysms suitable for treatment using double-branched device Bolton
排除标准
- •Failure to generate an adequate FE model of patient arteries (no preoperative multidetector contrast-enhanced CT-scan available, preoperative CT-scan slice thickness greater than 1mm, preoperative CT- scan with artifacts)
- •dissected aneurysm
结局指标
主要结局
In vitro versus stimulation test
时间窗: Day 0
To compare Vascutek in vitro validation test to the simulation of this test, the proportion of fLenestrations where \|L3 - L2\| and \|C3 - C2\| are less or equal to 2.5mm will be analysed. L3=distal edge of celiac trunk (L) after first simulation, stent-graft deployment in rigid aorta L2=distal edge of celiac trunk (L) after final custom stent-graft design, after in vitro tests
次要结局
未报告次要终点
