NCT04446247已完成不适用
Accuracy of Owlet Smart Sock Sensor 3rd Generation With Profound Hypoxia
适应症
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 14
- 试验地点
- 1
- 主要终点
- SpO2 Accuracy (Arms) of ≤3 Between 70 and 100%
研究概览
简要总结
Performance testing of pulse oximeters. The protocol involves brief stable arterial oxygen desaturation in healthy volunteers and sampling arterial blood when a stable level of hypoxia has been attained. The blood sample is analyzed for oxygen saturation with a gold standard bench CO-oximeter. This protocol is aligned with the latest ISO and FDA guidance documents for pulse oximeter testing.
研究设计
- 研究类型
- Observational
- 观察模型
- Other
- 时间视角
- Prospective
入排标准
- 年龄范围
- 18 Years 至 49 Years(Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •The subject is male or female, aged ≥18 and <
- •The subject is in good general health with no evidence of any medical problems.
- •The subject is fluent in both written and spoken English.
- •The subject has provided informed consent and is willing to comply with the study procedures.
排除标准
- •The subject is obese (BMI>30).
- •The subject has a known history of heart disease, lung disease, kidney or liver disease.
- •Diagnosis of asthma, sleep apnea, or use of CPAP.
- •Subject has diabetes.
- •Subject has a clotting disorder.
- •The subject a hemoglobinopathy or history of anemia, per subject report or the first blood sample, that in the opinion of the investigator, would make them unsuitable for study participation.
- •The subject has any other serious systemic illness.
- •The subject is a current smoker.
- •Any injury, deformity, or abnormality at the sensor sites that in the opinion of the investigators' would interfere with the sensors working correctly.
- •The subject has a history of fainting or vasovagal response.
- •The subject has a history of sensitivity to local anesthesia.
- •The subject has a diagnosis of Raynaud's disease.
- •The subject has unacceptable collateral circulation based on exam by the investigator (Allen's test).
- •The subject is pregnant, lactating or trying to get pregnant.
- •The subject is unable or unwilling to provide informed consent, or is unable or unwilling to comply with study procedures.
- •The subject has any other condition, which in the opinion of the investigators' would make them unsuitable for the study.
结局指标
主要结局
SpO2 Accuracy (Arms) of ≤3 Between 70 and 100%
时间窗: 1 hour
Determination of Average Root Mean Square SpO2 Accuracy of the Owlet Smart Sock 3rd Generation
次要结局
未报告次要终点
研究者
研究点 (1)
Loading locations...
相似试验
已完成
不适用
Accuracy of ScanWatch Pulse Oximeter With Profound HypoxiaOximetryNCT04380389Withings14
已完成
不适用
Owlet Smart Sock 3 Accuracy on InfantsHypoxia NeonatalNCT05635513Owlet Baby Care, Inc.28
已完成
不适用
SpO2 Accuracy Validation of the OxySoft Sensor Via Reference CO-Oximetry Motion StudyHypoxiaDesaturation of BloodNCT04559763Medtronic - MITG16
招募中
不适用
Evaluation of Accuracy of the Owlet OSS 3.0 Sensor in the Neonatal PopulationPulse Oximeter ValidationNCT06771206Owlet Baby Care, Inc.20
已完成
不适用
Accuracy of BabySat V3 With Profound Hypoxia and MotionHypoxiaNCT05581394Owlet Baby Care, Inc.12
