Influence of Different Toric Intraocular Lenses on IOL-capsular Complex After Cataract Surgery
试验速览
- 阶段
- 不适用
- 入组人数
- 100
- 主要终点
- The position of IOL
研究概览
简要总结
Corneal astigmatism is refractive error that impairs uncorrected visual acuity. When patients undergo cataract surgery, implantation of toric IOL is deemed the most effective choice for correcting corneal astigmatism and reducing postoperative spectacle dependence. The IOL-capsular complex is formed after cataract surgery and intraocular lens (IOL) implantation. Early postoperative mechanical wrapping of the anterior and posterior capsules plays a significant role in preventing IOL decentration and tilt, as well as formation of the IOL-capsular complex which reduces the incidence of posterior cataract opacity. IOL decentration ≥ 0.4 mm or/and IOL tilt ≥7degree were considered clinically significant because of poor post-surgery visual quality especially for optical sophisticated IOLs such as toric IOLs. IOL instability could decrease the curative effect of toric IOL. There are numerous types of toric IOLs; however, AcrySof IQ and TECNIS toric IOL are most frequently used in the Ophthalmology Hospital of Wenzhou Medical University. AcrySof IQ IOL has some differences compared with TECNIS toric IOL. IQ has a modified L haptic design while TECNIS® toric IOL has a modified C haptic design. AcrySof IQ IOL has a discontinuous 360˚ posterior square edge while TECNIS toric IOL has a continuous 360˚ posterior square edge. The AcrySof IQ IOL cylinder power models include 1.00, 1.50, 2.25, 3.00, 3.75, 4.50, 5.25 and 6.00 D. The Tecnis IOL cylinder power models include 1.00, 1.50, 2.25, 3.00 and 4.00 D. Previous study pointed that patients with cataracts with corneal astigmatism achieved comparable improvement in visual acuity, astigmatism correction, CS(sensitive contrast), rotational stability and satisfaction, following AcrySof and TECNIS toric IOL implantation. However,some other studies found that the Acrysof toric IOL showed significantly greater rotational stability than the Tecnis toric IOL and the rate of surgical IOL repositioning was higher in eyes implanted with TECNIS than with AcrySof toric IOLs for astigmatic correction. Currently, there is no literature guidance to compare the results of cataract surgery combined various types of toric IOLs implantation in patients using CASIA2. The novel anterior segment optical coherence tomography (AS-OCT) device, CASIA2 can accurately evaluate the IOL capsule bending and the lens position after cataract surgery. Also, CASIA2 can be used to documented the dynamic changes of IOL-capsular complex after surgery.
详细描述
Corneal astigmatism is refractive error that impairs uncorrected visual acuity. When patients undergo cataract surgery, implantation of toric IOL is deemed the most effective choice for correcting corneal astigmatism and reducing postoperative spectacle dependence. The IOL-capsular complex is formed after cataract surgery and intraocular lens (IOL) implantation. Early postoperative mechanical wrapping of the anterior and posterior capsules plays a significant role in preventing IOL decentration and tilt, as well as formation of the IOL-capsular complex which reduces the incidence of posterior cataract opacity.
Although up to 2-3 degree tilt and a 0.2-0.3 mm decentration are common and clinically unnoticed for any design of IOL, larger extent of tilt and decentration has a negative impact on the optical performance. IOL decentration ≥ 0.4 mm or/and IOL tilt ≥7degree were considered clinically significant because of poor post-surgery visual quality especially for optical sophisticated IOLs such as toric IOLs. IOL instability could decrease the curative effect of toric IOL. There are numerous types of toric IOLs; however, AcrySof IQ and TECNIS toric IOL are most frequently used in the Ophthalmology Hospital of Wenzhou Medical University. AcrySof IQ IOL has some differences compared with TECNIS toric IOL. IQ has a modified L haptic design while TECNIS® toric IOL has a modified C haptic design. AcrySof IQ IOL has a discontinuous 360˚ posterior square edge while TECNIS toric IOL has a continuous 360˚ posterior square edge. The AcrySof IQ IOL cylinder power models include 1.00, 1.50, 2.25, 3.00, 3.75, 4.50, 5.25 and 6.00 D. The Tecnis IOL cylinder power models include 1.00, 1.50, 2.25, 3.00 and 4.00 D. Previous study pointed that patients with cataracts with corneal astigmatism achieved comparable improvement in visual acuity, astigmatism correction, CS(sensitive contrast), rotational stability and satisfaction, following AcrySof and TECNIS toric IOL implantation. However,some other studies found that the Acrysof toric IOL showed significantly greater rotational stability than the Tecnis toric IOL and the rate of surgical IOL repositioning was higher in eyes implanted with TECNIS than with AcrySof toric IOLs for astigmatic correction. Currently, there is no literature guidance to compare the results of cataract surgery combined various types of toric IOLs implantation in patients using CASIA2. The novel anterior segment optical coherence tomography (AS-OCT) device, CASIA2 can accurately evaluate the IOL capsule bending and the lens position after cataract surgery. Also, CASIA2 can be used to documented the dynamic changes of IOL-capsular complex after surgery.
研究设计
- 研究类型
- Interventional
- 分配方式
- Randomized
- 干预模型
- Parallel
- 主要目的
- Treatment
- 盲法
- Quadruple (Participant, Care Provider, Investigator, Outcomes Assessor)
入排标准
- 年龄范围
- 50 Years 至 90 Years(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •The patients are diagnosed age related cataract or complicated with corneal astigmatism
- •The patients' age over 50 years old
- •The patients plan to receive cataract surgery in Eye hospital of Wenzhou Medical University
- •The dilated pupils are over 7mm
- •Patients are willing and able to complete the follow-ups.
排除标准
- •Patients with other type of cataract
- •Patients have complications in the surgery and after surgery
- •Patients have other severe diseases of eyes
研究组 & 干预措施
AcrySof IQ toric IOL
patients who had cataract surgery with AcrySof IQ toric IOL implantation. The patients diagnosed age related cataract The patients' ages are over 50.
干预措施: Cataract surgery with AcrySof IQ IOL implantation. (Procedure)
TECNIS toric IOL
patients who had cataract surgery with TECNIS toric IOL implantation. The patients are diagnosed age related cataract. The patients' ages are over 50.
干预措施: Cataract surgery with TECNIS toric IOL implantation. (Procedure)
结局指标
主要结局
The position of IOL
时间窗: The 3rd month after surgery
Evaluation of the position of IOL through tilt, decentration,rotation and the process of capsule bending using CAISA2(The novel anterior segment optical coherence tomography (AS-OCT) device)
The position of IOL
时间窗: The 1st day after surgery
Evaluation of the position of IOL through tilt, decentration,rotation and the process of capsule bending using CAISA2(The novel anterior segment optical coherence tomography (AS-OCT) device)
The position of IOL
时间窗: The 1st week after surgery
Evaluation of the position of IOL through tilt, decentration,rotation and the process of capsule bending using CAISA2(The novel anterior segment optical coherence tomography (AS-OCT) device)
The position of IOL
时间窗: The 1st month after surgery
Evaluation of the position of IOL through tilt, decentration,rotation and the process of capsule bending using CAISA2(The novel anterior segment optical coherence tomography (AS-OCT) device)
次要结局
未报告次要终点
研究者
Yune Zhao
Vice president of Eye Hospital of Wenzhou Medical University
Wenzhou Medical University