跳至主要内容
临床试验/NCT00180479
NCT00180479已完成3 期

SPIRIT III: A Clinical Evaluation of the Investigational Device XIENCE V® Everolimus Eluting Coronary Stent System (EECSS) in the Treatment of Subjects With de Novo Native Coronary Artery Lesions

Abbott Medical Devices128 个研究点 分布在 1 个国家目标入组 1,002 人开始时间: 2005年6月最近更新:
适应症

试验速览

阶段
3 期
状态
已完成
入组人数
1,002
试验地点
128
主要终点
Primary Endpoint: In-segment Late Loss (LL)

研究概览

简要总结

This study is divided into 5 arms:

  1. Randomized Clinical Trial (RCT): Prospective, randomized, active-controlled, single blind, parallel two-arm multi-center clinical trial in the United States (US) comparing XIENCE V® Everolimus Eluting Coronary Stent System (CSS) (2.5, 3.0, 3.5 mm diameter stents) to the Food and Drug Administration (FDA) approved commercially available active control TAXUS® EXPRESS2™ Paclitaxel Eluting Coronary Stent (TAXUS® EXPRESS2™ PECS) System
  2. US 2.25 mm non-randomized arm using 2.25 mm diameter XIENCE V® Everolimus Eluting CSS
  3. US 4.0 mm non-randomized arm using 4.0 mm diameter XIENCE V® Everolimus Eluting CSS
  4. US 38 mm non-randomized arm using 38 mm in length XIENCE V® Everolimus Eluting CSS
  5. Japanese non-randomized arm using XIENCE V® Everolimus Eluting CSS (2.5, 3.0, 3.5, 4.0 mm diameter stents) in Japan

The TAXUS® EXPRESS2™ Paclitaxel Eluting Coronary Stent System is Manufactured by Boston Scientific.

详细描述

The purpose of the SPIRIT III clinical trial is to evaluate the safety and efficacy of the XIENCE V® Everolimus Eluting Coronary Stent System (XIENCE V® EECSS). The XIENCE V® EECS (XIENCE V® arm) will be compared to an active control group represented by the FDA approved commercially available Boston Scientific TAXUS® EXPRESS2™ Paclitaxel-Eluting Coronary Stent (TAXUS® EXPRESS2™ PECS) System (TAXUS® arm).

The SPIRIT III clinical trial consists of a randomized clinical trial (RCT) in the US which will enroll approximately 1,002 subjects (2:1 randomization XIENCE V® EECS : TAXUS® EXPRESS2™ PECS) with a maximum of two de novo native coronary artery lesion treatment within vessel sizes >= 2.5 mm and <= 3.75 mm.

The SPIRIT III clinical trial also consists of three concurrent US non-randomized arms (2.25 mm diameter stent, 4.0 mm diameter stent and 38 mm length stent arms) and one Japanese non-randomized arm as follows:

  1. 105 subjects with a maximum of two de novo native coronary artery lesion within vessel sizes > 2.25 mm and < 2.5 mm and lesion length <= 22 mm will be enrolled concurrently in the US 2.25 mm non-randomized treatment arm
  2. 80 subjects with a maximum of two de novo native coronary artery lesion within vessel sizes > 3.75 mm and >= 4.25 mm and lesion length <= 28 mm will be enrolled concurrently in the US 4.0 mm non-randomized treatment arm
  3. 105 subjects with a maximum of two de novo native coronary artery lesion within vessel sizes > 3.0 mm and < 4.25 mm and lesion length > 24 mm and < 32 mm will be enrolled concurrently in the US 38 mm non-randomized treatment arm.
  4. 88 Japanese subjects with a maximum of two de novo native coronary artery lesions within vessel sizes >= 2.5 mm and <= 4.25 mm and lesion length <= 28 mm will be enrolled concurrently in the non-randomized Japanese arm.

All subjects in the RCT and the four non-randomized arms will be screened per the protocol required inclusion/exclusion criteria. The data collected will be compared to data from the subjects enrolled into the TAXUS® arm of US RCT.

研究设计

研究类型
Interventional
分配方式
Randomized
干预模型
Parallel
主要目的
Treatment
盲法
Single (Participant)

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Target lesion(s) must be located in a native epicardial vessel with visually estimated diameter between >= 2.25 mm and <= 4.25 mm and a lesion length <= 32 mm
  • The target lesion(s) must be in a major artery or branch with a visually estimated stenosis of >= 50% and < 100% with a thrombolysis in myocardial infarction (TIMI) flow of >= 1
  • Non-study, percutaneous intervention for lesions in a non-target vessel is allowed if done >= 90 days prior to the index procedure (subjects who received brachytherapy will be excluded from the trial)

排除标准

  • Located within an arterial or saphenous vein graft or distal to a diseased (vessel irregularity per angiogram and > 20% stenosed lesion by visual estimation) arterial or saphenous vein graft
  • Lesion involving a bifurcation >= 2 mm in diameter or ostial lesion > 50% stenosed by visual estimation or side branch requiring predilatation
  • Located in a major epicardial vessel that has been previously treated with brachytherapy
  • Located in a major epicardial vessel that has been previously treated with percutaneous intervention < 9 months prior to index procedure
  • Total occlusion (TIMI flow 0), prior to wire passing
  • The target vessel contains thrombus
  • Another significant lesion (> 40% diameter stenosis [DS]) is located in the same epicardial vessel as the target lesion

结局指标

主要结局

Primary Endpoint: In-segment Late Loss (LL)

时间窗: 240 days

In-segment minimal lumen diameter (MLD) post-procedure minus (-) in segment MLD at 240 day follow-up and 5 mm proximal and 5mm distal to the stent equals Late Loss. MLD defined: The average of two orthogonal views (when possible) of the narrowest point within the area of assessment.

次要结局

  • Major Secondary Endpoint: Ischemia Driven Target Vessel Failure (ID-TVF)(270 days)
  • Target Vessel Failure (TVF)(5 years)
  • Ischemia Driven Target Lesion Revascularization (ID-TLR)(5 years)
  • Ischemia Driven Target Vessel Revascularization (ID-TVR)(5 years)
  • Ischemia Driven Major Adverse Cardiac Event (MACE)(5 years)
  • Ischemia Driven Major Adverse Cardiac Event(MACE)(2 years)
  • In-stent % Angiographic Binary Restenosis (% ABR) Rate(at 240 days)
  • In-segment % Angiographic Binary Restenosis (% ABR) Rate(240 days)
  • Persisting Incomplete Stent Apposition, Late-acquired Incomplete Stent Apposition, Aneurysm, Thrombosis, and Persisting Dissection(at 240 days)
  • Acute Success: Clinical Device(In-hospital)
  • Acute Success: Clinical Procedure(In-hospital)
  • Proximal Late Loss(at 240 days)
  • Distal Late Loss(240 days)
  • In-stent Late Loss(at 240 days)
  • % Volume Obstruction (% VO)(at 240 days)
  • In-stent % Diameter Stenosis (% DS)(at 240 days)
  • In-segment % Diameter Stenosis (% DS)(240 days)

研究者

申办方类型
Industry
责任方
Sponsor

研究点 (128)

Loading locations...

相似试验