跳至主要内容
临床试验/NCT04618848
NCT04618848Unknown不适用

Performance Evaluation Protocol for Sight Diagnostics' OLO CBC Analyzer by Ichilov Oncology Daily Care Center

Sight Diagnostics0 个研究点目标入组 50 人开始时间: 2020年11月最近更新:
适应症

试验速览

阶段
不适用
发起方
入组人数
50
主要终点
CBC Comparison

研究概览

简要总结

This study aims to establish OLO's performance accuracy of CBC results for finger prick samples across the OLO reportable range, and particularly around relevant medical decision points

详细描述

This study aims to establish OLO's performance accuracy of CBC results for finger prick samples across the OLO reportable range, and particularly around relevant medical decision points.

The evaluation will include two studies:

  1. A matrix study will compare a capillary sample collected via finger prick with a venous sample collected into a blood tube, both scanned on OLO.
  2. A method comparison study will compare Sight OLO to the site reference analyzer using a venous sample collected into a blood tube.

研究设计

研究类型
Observational
观察模型
Case Only
时间视角
Cross Sectional

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Blood samples will be collected from oncologic patients older than 18 years old arriving at the oncologic daily care center of Ichilov hospital, with predefined normal and abnormal blood counts

排除标准

  • Patient younger than 18 years old
  • Visibly hemolyzed specimens (sample plasma should be checked for hemolysis by technician while selection of specimen from tube racks).
  • Visibly clotted specimens (Reasonable to assume that samples processed with clinical reported results are not clotted).
  • A sample that is not successfully scanned on OLO within 3 attempts (e.g. rejected due to irregular cell morphology, or microscopically visible hemolysis).

结局指标

主要结局

CBC Comparison

时间窗: Immediately after sample collection is completed

Comparison between CBC results obtained from the Sight OLO and the reference device in venous and capillary samples covering medical decision points relevant to onclology patient

次要结局

未报告次要终点

研究者

发起方
Sight Diagnostics
申办方类型
Industry
责任方
Sponsor

相似试验